Operator's Manual

70
RefeRences and TRoubleshooTing
RefeRences and
TRoubleshooTing
Specifications (Nominal)
General Pump Specifications
System definition CADD®-Solis pump with 1 of the following attached:
Medication cassette reservoir and CADD® extension set
Medication cassette reservoir with Flow Stop feature and CADD® extension set
CADD®
administration set
CADD
®
administration set with Flow Stop feature
Classification
CF J
Class II K
Used to test the
pump
CADD
medication cassette reservoirs, < 21-7002 and < 21-7309
CADD® extension sets, < 21-7047 and < 21-7046
CADD® administration sets, < 21-7091 and < 21-7321
CADD® high volume administration sets, < 21-7355 and < 21-7357
Resolution CADD
medication cassette reservoir: 0.05 mL per pump stroke nominal
CADD® administration set: 0.05 mL per pump stroke nominal
CADD®
high volume administration set: 0.1 mL per pump stroke nominal
Size Excluding cassette and accessories:
4.1 cm × 10.2 cm × 12.7 cm
1.6 in × 4 in × 5 in
Weight Including 4 AA alkaline batteries, excluding other accessories:
595 g
21 oz
System operating
temperature
2°C to 40°C
36°F to 104°F
System storage
and transportation
temperature
20°C to 60°C
4°F to 140°F
Maximum surface
Temperature
(pump, while
operating)
48°C (11F) Maximum
Moisture
protection
Splashproof ( E ) per IEC 60529
Relative humidity 20% to 90% relative humidity, non-condensing
Atmospheric
pressure
70 kPa to 106 kPa
10.2 psi to 15.4 psi
Expected Service
Life
5 years
Power sources AC adapter
CADD®-Solis rechargeable battery pack
CADD®-Solis Communication Module with integrated rechargeable battery
Four AA alkaline batteries (for example, Duracell
®
PC1500 / MN1500, IEC LR6)
Charging system
for internal
memory backup
battery
The internal memory backup battery uses lithium manganese dioxide
technology. It charges whenever the pump is powered on and has a 10-month
memory capacity once it has been charged for 250 hours at 20°C (68°F).
Battery fallout
alarm
Alarm sounds for at least 2 minutes if the pump has been powered up for a
minimum of 4 minutes. Note: Alarm enabled while pump is in run mode only.