Operator's Manual
63
RefeRences and TRoubleshooTing
RefeRences and
TRoubleshooTing
Alarms and
Messages
Alarm
Priority
Description / Corrective Action
Protocol library
updating.
Reselect protocol
when update is
complete.
Medium A new or updated library is currently being downloaded into the
pump. You cannot select any new protocols while this update
is in process. Select Acknowledge to clear the alarm or it will
automatically clear when the update is complete.
Pump does not
have a protocol
library.
Medium The pump had a protocol library the last time it was powered on,
but now it does not. This could happen if the pump was manually
reverted to the factory default, recently had a software update, or if
an attempt to install a protocol library failed. Select Acknowledge
to clear the alarm and refer to your facility’s PharmGuard® system
administrator to download a new protocol library.
Pump settings and
patient data lost.
Medium The pump reverted to the factory default. The pump was manually
reverted to this default, has recently had a software update, or has
not been in use for some time. Select Acknowledge to clear the
alarm and refer to your facility’s PharmGuard® system administrator
to reprogram the pump.
Pump stopped by
an alarm that has
since cleared.
High The pump was stopped by another high priority alarm. That
alarm was not acknowledged, but the problem has since cleared.
Select Acknowledge to clear the alarm and restart the pump, if
appropriate. The event log has recorded the alarm that stopped the
pump. For information on accessing the event log, see page 43.
Pump stopped
reminder. When
ready, press
‘stop/start’ key.
High This is a reminder that the pump is stopped and not delivering.
Select Acknowledge to clear the alarm. Start the pump, if
appropriate. The alarm will repeat in 5 minutes if the pump is not
restarted or powered down.
Rechargeable
battery near end
of use. Replace
battery.
Medium The rechargeable battery pack or the CADD®-Solis Communication
Module is near the end of its life. It has been discharged and
recharged so many times that it will soon be at the end of its use.
You may continue to use it in this state.
Rechargeable
battery reached
end of use. Pump
will not run.
High The rechargeable battery pack or the CADD®-Solis Communication
Module is at the end of its life. It has been discharged and recharged
so many times that it is no longer able to hold a good charge.
Remove the battery from service. Install 4 new AA batteries or a fully
charged rechargeable battery pack. To start delivery, good batteries
must always be installed even when an external source of power is
connected.
Remove and
reattach cassette.
High The pump detects a damaged cassette. Close the tubing clamp and
inspect the cassette for damage. The pump is stopped and will not
run. Replace the cassette if necessary.
Note: You must remove the cassette to continue.
Note: This alarm also occurs if a cassette is attached during one of
the following situations:
• The pump is set to the factory default, powered off, and then
powered on.
• The pump was loaded with new software, powered off, and then
powered on.
Reservoir volume
is zero. Pump
stopped.
High The reservoir volume has reached 0.0 mL. The pump was delivering,
is now stopped, and will not run. Select Acknowledge to clear the
alarm. Install a new fluid container and edit the value of the reservoir
volume, if appropriate.