Operator's Manual

59
RefeRences and TRoubleshooTing
RefeRences and
TRoubleshooTing
Alarms and
Messages
Alarm
Priority
Description / Corrective Action
Battery removed.
Pump stopped.
High The rechargeable battery pack, CADD®-Solis Communication
Module with integrated rechargeable battery, or the 4 AA batteries
were removed. The pump was delivering and is now stopped. Select
Acknowledge to clear the alarm.
Install 4 new AA batteries, a fully charged rechargeable battery
pack, or CADD®-Solis communication module. To start delivery,
good batteries must always be installed, even when an external
source of power is connected.
If appropriate, restart the pump.
Battery removed.
Pump will not run.
Medium The pump is stopped and the rechargeable battery pack,
CADD®-Solis Communication Module with integrated rechargeable
battery, or the 4 AA batteries were removed, but the pump is still
powered by the AC adapter. Select Acknowledge to clear the alarm.
Install a fully charged rechargeable battery pack, CADD®-Solis
communication module, or 4 new AA batteries. To start delivery,
good batteries must always be installed, even when an external
source of power is connected.
Cannot
start pump.
Rechargeable
battery has
reached end of
use.
Medium The rechargeable battery pack or the rechargable battery integrated
with the CADD-Solis Network Comunication Module is at the end of
its life. It has been discharged and recharged so many times that it
is no longer able to hold a good charge. Remove the battery from
service.
To start delivery, good batteries must always be installed even when
an external source of power is connected. Select Acknowledge
to clear the alarm. Install 4 new AA batteries, a new rechargeable
battery pack, or the CADD®-Solis Communication Module with a its
rechargeable battery replaced. If appropriate, start the pump.
Cannot start pump
with a depleted
battery.
Medium The battery power is too low to operate the pump. To start delivery,
good batteries must always be installed even when an external
source of power is connected. Select Acknowledge to clear the
alarm. Install 4 new AA batteries, a fully charged rechargeable
battery pack, or attach an AC adapter to charge a rechargeable
battery pack or communication module battery. If appropriate, start
the pump.
Cannot start pump
with a reservoir
volume of zero.
Medium The reservoir volume in the pump is set to zero. Select acknowledge
to clear the alarm. Edit the reservoir volume to the correct value.
If appropriate, start the pump. See Reservoir Volume on page 50 for
more information about the reservoir volume.
Cannot start the
pump with air
in-line. Prime
tubing.
Medium The air detector has detected air in the fluid path directly under
the air detector; the fluid path may contain air bubbles. Select
Acknowledge to clear the alarm, then:
If the fluid path contains air bubbles, close the clamps and
disconnect the fluid path from the patient. Follow the instructions
for removing air by priming (see page 34 ).
If appropriate, start the pump.
Cannot start pump
with an unusable
battery
Medium The batteries installed are either the wrong kind of AA batteries, or
you are using a rechargeable battery pack that is not compatible
with the pump. To start delivery, good batteries must always be
installed even when an external source of power is connected. Select
Acknowledge to clear the alarm. Remove the batteries and install
the 4 new AA batteries, or fully-charged rechargeable battery pack.
If appropriate, start the pump.
For more information on what type of batteries to use, see Installing
the Batteries on page 20