Operator's Manual
58
RefeRences and TRoubleshooTing
RefeRences and
TRoubleshooTing
Alarms and Messages, Alphabetical List
Alarms and
Messages
Alarm
Priority
Description / Corrective Action
(Screen is blank
and alarm is
sounding)
High The pump was delivering and the batteries or the CADD®-Solis
Communication Module with integrated rechargeable battery
was removed, the battery door was opened, or the batteries are
too depleated to power the pump. The pump lost power and is no
longer delivering. Clear this alarm by turning the pump back on, or
the alarm stops after the power is off for a minimum of 2 minutes.
A setting was
edited, but not
saved, and the edit
was lost.
Medium A parameter was being manually edited, but it wasn’t saved and the
pump reverted to the Home screen. Select Acknowledge to clear
the alarm, and if appropriate, edit the parameter and save.
AC Adapter
disconnected.
Low The AC adapter was disconnected and the pump is being powered
by the 4 AA batteries , rechargeable battery pack or the CADD®-Solis
Communication Module with integrated rechargeable battery. Select
Acknowledge to clear the alarm or the alarm will automatically clear
after 5 seconds. If desired, reconnect the AC adapter.
Air in-line
detected. Press
‘Acknowledge’
then prime tubing.
High The air detector has detected accumulated air in the fluid path; the
fluid path may contain air bubbles. The pump was delivering and
is now stopped and will not run. Select Acknowledge to clear the
alarm.
If the fluid path contains air bubbles, close the clamps and
disconnect the fluid path from the patient. Follow the instructions
for removing air by priming (see Prime Tubing on page 34 for more
information on priming). Restart the pump.
Battery depleted.
Pump stopped.
High The pump was delivering and is now stopped and the battery power
is too low to operate the pump. If the AC adapter is attached, select
Acknowledge to clear the alarm. If the AC adapter is not attached,
select Power Down to power down the pump.
• Charge the rechargeable battery pack or the CADD®-Solis
Communication Module with integrated rechargeable battery.
• Install 4 new AA batteries or a fully charged rechargeable battery
pack. To start delivery, good batteries must always be installed,
even when an external source of power is connected.
• If appropriate, restart the pump.
Battery depleted.
Pump will not run.
Medium The battery power is too low to operate the pump. If the AC adapter
is attached, select Acknowledge to clear the alarm. If the AC adapter
is not attached, select Power Down to power down the pump.
• Charge the rechargeable battery pack or the CADD®-Solis
Communication Module with integrated rechargeable battery.
• Install 4 new AA batteries or a fully charged rechargeable battery
pack. To start delivery, good batteries must always be installed,
even when an external source of power is connected.
Battery low.
Replace battery.
Low The rechargeable battery pack, CADD®-Solis Communication Module
with integrated rechargeable battery, or 4 AA batteries are low but
the pump is still operable. Select Acknowledge to clear the alarm or
the alarm automatically clears after 5 seconds. Recharge or change
the rechargeable battery pack or CADD®-Solis communication
module, or replace the 4 AA batteries soon.