Operator's Manual

53
Programming and oPeration
Programming
and oPeration
To use the manual mode:
1. Determine whether you are starting a new patient, or starting a new protocol with the same patient.
Select the appropriate task from the tasks menu.
2. On the Select Therapy screen, scroll to the bottom of the therapies until
[Program Manually] is highlighted and press S. The protocol title bar
is displayed in the color black. Instead of a therapy and qualifier, you see the
Manual Program screen (see page 51).
3. Unlock the keypad using the security code.
4. Press U or D to select the desired units (mL, mg, mcg), and then press
S.
If you selected mg or mcg, press U or D to scroll to the desired
concentration and select Confirm. If you selected mL, proceed to the next
step.
5. The therapy, qualifier, units and concentration, if applicable, that you have
selected now appear. Confirm that you have selected the correct units and
concentration and select Yes.
Note: If the information displayed is incorrect, selecting the left soft key
backs you out of each screen, allowing you to start over.
6. The pump takes a moment to set the program before it asks you to review
the pump settings. Select Review.
7. Carefully review each patient specific parameter.
If the parameters are not at the desired values, press S to edit. (See Patient Specific Parameters on page
45 for more information.)
If the parameters are correct, select Accept Value.
Note: When the pump is in the manual mode, only the factory default settings are considered to be
within the soft limits. If you edit any of the parameters outside of the factory default, you will exceed
the soft limit range and will be asked to confirm the soft limit override on each screen (see page 82
for the factory default settings). The parameter and value are displayed in amber.
WARNING: The manual mode does not contain programming limits. Carefully review each parameter to
ensure it accurately matches the prescription. Failure to set the manual mode to the correct values
could result in serious patient injury or death.
8. Once you have accepted each of the values, follow the instructions on the pump to attach the cassette,
prime the set, and start the pump.