Manual Mode Initial Settings
Pump Function Protocol Setting
Programming Units User selects mL, mg, or mcg
Concentration User selects if the programming units are mg or mcg.
Range is 0.1100 mg/mL or 1500 mcg/mL
Continuous Rate On/Off On
Continuous Rate Limits Range is 0100 mL/hour (or mg or mcg equivalent).
Value is 0
PCA Dose On/Off On
PCA Dose Limits Range is 0–50 mL (or mg or mcg equivalent).
Value is 0
PCA Lockout Limits Range is 1 min–24 hours.
Value is 1 hour (when PCA dose is programmed)
Intermittent Bolus On/Off Off
Intermittent Bolus Limits Range is 0–50 mL.
Value is 0 mL
Bolus Interval Limits Range is 1 minute–4 hours.
Value is 1 hour
Intermittent Bolus Type Bolus Interval or PCA Lockout.
Value is Bolus Interval
Next Bolus Range is 0 minutes–4 hours.
Value is 0 hr
Delivery limit amount* Range is 0.11000 mL (or mg or mcg equivalent)
Max doses/hour* Range is 1–60.
Value is 1
Reservoir volume 1 mL
Reservoir volume reset value 100 mL
Delivery limit method Remains the same as the previously used protocol
Reservoir volume low trip point 5.0 mL
Reservoir low alarm type Remains the same as the previously used protocol
Clinician bolus amount Range is 0–50 mL (or mg or mcg equivalent)
Delivery limit period Remains the same as the previously used protocol
Maximum Delivery Rate Remains the same as the previously used protocol
Pump stopped alarm type Remains the same as the previously used protocol
Upstream sensor on/off Remains the same as the previously used protocol
Downstream sensor sensitivity Remains the same as the previously used protocol
Air detector on/off On
Air detector sensitivity Remains the same as the previously used protocol
Time and date settings Remains the same as the previously used protocol
PM reminder on/off Remains the same as the previously used protocol
PM reminder interval Remains the same as the previously used protocol
Display and sound settings Remains the same as the previously used protocol
Security settings Remains the same as the previously used protocol
*Only if feature was used in the previous protocol.