Programming and oPeration
Programming
and oPeration
Manual Mode Programming
The manual mode is for situations when the doctor’s orders may not match any of the protocols in the library,
or when a protocol library is not available. The screen for the manual mode is always black in color and the
therapy, qualifier, and drug and concentration listing will look like this (see page 23 for a diagram of the
pump screen):
Note: The protocol title bar may be black even when the pump is not in manual mode if the protocol was
modified in the administrator settings (see the Administrator Settings Guide for more information). Be
sure to check the therapy, qualifier, drug, and concentration or units displayed to ensure you are in manual
mode.
Unlike protocols that are created and downloaded into the pump by the PharmGuard® Medication Safety
Software System, the manual mode allows you to choose the units (mL, mg, mcg) and concentration, and
does not contain any programming limits. Many of the parameters in the manual mode remain the same as
the parameters from the previously used protocol. For example, if the previous protocol had delivery limit
set as its delivery limit method, the manual mode will also have delivery limit as its delivery limit method.
However, the actual programming limits of the previous protocol are erased. The hard and soft limits are all
set to the factory default, which means the delivery ranges are not limited. See the following chart for details.
Note: If the pump was set to the factory default, or if you are using the pump for the first time, the previously
selected protocol is the factory default settings. Refer to the table on page 82 to see what the factory
default settings are.
If you desire to set programming limits after choosing the manual mode, you may do so. See the
Administrator Settings Guide for more information.
Note: The manual mode is designed to be used in unusual circumstances when an order is received which is
different than any of your standard protocols. Use of the manual mode does not allow you to use all of the
medication safety features built into the CADD
®
-Solis pump. If you find the manual mode is being used
frequently, consider contacting your PharmGuard® system administrator to discuss if additional protocols
should be created.
Units and concentration chosen after the
manual mode is selected.