Operator's Manual
46
Programming and oPeration
Programming
and oPeration
Intermittent Bolus
Each intermittent bolus is the infusion of a specific volume of drug at the programmed max delivery rate
value. Programming the intermittent bolus is limited by the values of the max delivery rate, bolus interval,
and continuous rate. Intermittent boluses are delivered at regular time intervals based on the programmed
Bolus Interval, and as allowed by intermittent bolus limits, bolus interval limits, and bolus type. For more
information, contact your PharmGuard® system administrator.
This value can be edited while the pump is running or stopped (if the keypad is unlocked).
Note: Stopping the pump while an Intermittent Bolus is in progress will cancel the remainder of the
Intermittent Bolus dose.
To edit the intermittent bolus:
1. From the Home screen, press U or D to choose Intermittent Bolus and
press S.
2. Unlock the keypad using the security code or the pump key.
3. Press U or D until you reach the desired intermittent bolus quantity
and select Save.
Bolus Interval
The bolus interval is the amount of time that elapses from the beginning of one intermittent bolus to the
beginning of the next one. The bolus interval cannot be programmed to a value that would prevent an
intermittent bolus from being fully delivered within the bolus interval time period. Programming the bolus
interval is limited by values of the max delivery rate, continuous rate, and intermittent bolus. For more
information, contact your PharmGuard® system administrator.
This value can be edited while the pump is running or stopped (if the keypad is unlocked).
To edit the bolus interval:
1. From the Home screen, press U or D to choose Bolus Interval and press
S.
2. Unlock the keypad using the security code or the pump key.
3. Press U or D until you reach the desired bolus interval and select Save.