Operator's Manual
42
Programming and oPeration
Programming
and oPeration
Current Profile / Protocol (if configured)
The Current Profile / Protocol screen shows the profile, therapy, qualifier, drug
and units in use based on your selected protocol or manually programmed
protocol.
To view the current profile / protocol:
1. From the Home screen, select Reports.
2. From the Reports menu, press U or D to choose Current Profile /
Protocol and press S.
Delivery Log
The delivery log is a subset of the event log and contains information having to do specifically with delivery
events. Delivery log information includes:
•
PCA dose deliveries.
•
Clinician boluses.
•
Intermittent boluses.
•
Changes to the patient specific parameters (including continuous rate, intermittent boluses, PCA
dose, PCA dose lockout, delivery limit).
•
Manually stopping a PCA dose and/or a clinician bolus.
•
Starting a new protocol.
•
Pump started, stopped, powered up, and powered down.
The delivery log is maintained by the pump, and displays all entries since the last time a new patient was
started. The delivery log can be viewed at any time, with the pump running or stopped.
To view the delivery log:
1. From the Home screen, select Reports.
2. From the Reports menu, press U or D to choose Delivery Log and
press S.
While viewing the delivery log, you may
quickly scroll from the oldest to newest by
selecting Show Oldest or Show Newest.