Operator's Manual

33
Programming and oPeration
Programming
and oPeration
6. Press U or D to highlight the desired qualifier and press S.
7. Press U or D to highlight the desired drug(s) and concentration,
and press S.
8. Unlock the keypad using the security code.
9. The new protocol is displayed on the screen and the pump asks if you have
selected the correct therapy, qualifier, drug, and concentration.
Confirm that you have selected the correct profile, therapy, qualifier,
drug, and concentration or units.
Note: If programming in mL, you are not asked to confirm the concentration.
Select Yes to program the pump with the protocol you have chosen. Select
No to go back to the Select Drug screen. If you want to change the drug,
repeat the process for selecting a new drug. Otherwise, select Back until you reach the screen you want
to change (profile, therapy or qualifier). You will be asked to enter a security code after selecting Back and
reselecting any profile, therapy, qualifier, or drug.
10. You will be prompted to review the pump settings to accept each value before the pump can be started.
During the review, the pump settings may be edited if necessary.
Note: If a clinical advisory note has been defined by your facility for the selected protocol, it will appear
after accepting all values. To acknowledge the clinical advisory note, select OK. If you have a question
about the note, contact your PharmGuard® system administrator.
Start New Protocol, Same Patient
The process for starting a new protocol for the same patient is much like starting a new patient; however, the
event log does not insert a new patient marker (all other delivery related reports are cleared except the delivery
log). See Start New Patient on page 32 for directions on selecting a profile, therapy, qualifier, and drug for the
new protocol.
Note: When starting a new protocol, be sure to attach a new reservoir with the proper drug and concentration.