Operator's Manual

32
Programming and oPeration
Programming
and oPeration
4. Unlock the keypad using the pump key (if allowed) or press U or D to
enter the clinician code (or a higher level code). Press
S
to advance to
the next digit. Once the code is entered, select Accept Value.
Note: If you enter a code that you believe is correct and receive a wrong
code error, check the screen to see if the clinician code is customized.
If the code was not customized, it was entered incorrectly. Retry. If the
code is customized, use the custom clinician code. If you do not know
this code, contact your PharmGuard® system administrator.
WARNING: To prevent the patient from accessing the clinician bolus
function, do not let the patient know the security codes. Improper
programming could result in serious patient injury or death.
5. Make sure the clinician bolus amount is at the desired value and select
Deliver.
Note: If you enter a value outside the soft limit, a screen appears asking you
to confirm the soft limit override.
6. The screen shows the amount decreasing as the bolus is delivered. You may
stop the bolus at any time by selecting Stop Bolus.
WARNING: Never leave the pump unattended while on the clinician bolus
edit screen. You must deliver the programmed value, or cancel to leave
the screen. Failure to do so could result in serious patient injury or death.
Start New Patient
Each time a new patient is started, it is recorded in the event log. All other delivery related reports are
cleared.
To start a new patient:
1. Stop the pump if it is running.
2. From the Home screen, select Tasks.
3. From the Tasks menu, press U or D until Start New Patient is
highlighted and press S.
4. If your library was setup to use profiles by the PharmGuard® system
administrator, the previously selected profile is shown. Profiles contain a
grouping of protocols (e.g., profiles can be defined for specific care areas,
etc.). If your desired therapy is contained within the profile shown, select
Yes, otherwise select No. If No is selected, you may press U or D to
highlight the desired profile and press S.
5. The therapies contained within your selected profile are shown.
Press U or D to highlight the desired therapy and press S.