Operator's Manual

31
Programming and oPeration
Programming
and oPeration
Tasks
The tasks menu leads you to most of the pump’s operating functions. On this
menu, you can perform a range of tasks from giving a clinician bolus, starting a
new patient, to changing the date and time, turning wireless on/off, or adjusting
the PharmGuard® system administrator settings. Some of the items on the tasks
menu are protected by the various security levels. To learn more about the
security codes, see Security Settings on page 28. To access the tasks menu,
select Tasks on the Home screen.
The following functions can be found on the tasks menu:
Tasks
Give Clinician Bolus
Start New Patient
Start New Protocol, Same Patient
Prime Tubing
Set Time and Date
Adjust Backlight Intensity
Adjust Alarm Volume
View Reports
Turn Wireless On/Off
Adjust Admin Settings
Give Clinician Bolus
A clinician bolus may only be delivered while the pump is running. It allows you to deliver a specified
amount of drug, for example, as a loading dose. A clinician bolus cannot be started while a PCA (PCEA) dose
or intermittent dose is in progress. The amount delivered decreases the reservoir volume and increases
the given amount, but does not add to the dose counters or to the delivery limit. A clinician bolus may be
stopped in progress.
Note: If a clinician bolus is manually stopped by a clinician or automatically stopped by an alarm, power
failure, or other condition that stops delivery, the pump remembers what point the bolus was at when it
stopped. The next time you choose to give a clinician bolus, it gives you the option to restart the clinician
bolus where it left off or to start with a new clinician bolus.
WARNING: Exercise care when using the clinician bolus function. Because there are no limits on the frequency
of delivering a bolus, and because the bolus amount may be set as high as 50 mL (or the mg or mcg
equivalent), do not permit the patient or an unauthorized clinician to become familiar with the procedure
for giving a clinician bolus. Improper programming could result in serious patient injury or death.
Note: The maximum clinician bolus may be limited by the settings in the protocol, which is determined by the
PharmGuard® system administrator.
To give a clinician bolus:
1. Make sure the pump is running. Start the pump if necessary.
2. From the Home screen, select Tasks.
3. From the Tasks menu, press U or D until Give Clinician Bolus is
highlighted, then press S.