Programming and oPeration
Programming
and oPeration
Autolock
The CADD
®
-Solis ambulatory infusion pump is designed to meet both safety and usability needs. The
autolock feature reduces the chance of unauthorized pump programming. When the keypad is unlocked with
a security code and left unlocked, the software automatically locks the keypad.
The autolock feature takes affect on the Home screen approximately 30 seconds after the last key press, or
when reverting back to the Home screen from other screens. Autolock takes longer on programming or task
screens when you typically need more time to perform an action. From some of these screens, it may take up
to 4 minutes after the last key press to revert back to the Home screen. If the pump is alarming, autolock does
not take affect.
If using the key to unlock the keypad while the pump is running, you should relock the keypad using the key.
Note: The keypad can be re-locked by pushing the right soft key twice from the Home screen or once from
the tasks menu. As a recommended safety precaution, you should always manually lock the pump using
this feature.
Entering Security Codes
To enter a security code when prompted:
1. Press U or D to scroll up or down to the correct digit. Press S or
Accept Value to advance to the next digit. Once the complete code has been
entered, press
S
or Accept Value.
2. If an incorrect code is entered, a wrong code error appears. It includes the
security level required to access the function. Select Retry to enter the code
again.
Note: If you enter a code that you believe is correct and receive a wrong
code message, check the screen to see if the code has been customized.
If the code is not customized, it was entered incorrectly and you should
retry. If the code is customized, enter the customized code. If you do not
know the code, contact your PharmGuard® system administrator.
3. Depending on the level of security required and
how the PharmGuard® system administrator has
programmed the security settings, you may be
able to use the pump key to unlock the keypad.
When accessing a menu or function that requires
a security code, 1 of 2 prompts appears, as shown
in the sample screens to the right.
To unlock the keypad with the pump key, turn the pump key counterclockwise.
Note: It is possible for the keypad to be locked while the cassette/keypad
lock is unlocked. To use the pump key to unlock the keypad, first lock the
cassette/keypad lock, and then unlock it.