Operator's Manual

26
PumP SetuP
PumP SetuP
6. Lift the cassette latch into the closed position. You should be able to
move the latch into the closed position with minimal to no resistance.
If you experience resistance when lifting the cassette latch, DO NOT
FORCE the latch. If you are unable to attach the cassette to the pump
with minimal to no resistance, the cassette is not in the proper latching
position. If the pump does not latch easily, unlatch the cassette and
repeat the process. If unsuccessful on the second attempt, do not
use the pump. Contact Smiths Medical Customer Service for further
assistance.
7. Verify the cassette is attached correctly. Looking from left to right, the top of the cassette should line up
evenly with the bottom of the pump and be securely attached. If the cassette is attached incorrectly, there
will be an uneven gap between the cassette and the pump, with the gap appearing on the latch side of
the pump. If an uneven gap exists, unlatch the cassette and repeat the process. If the gap exists after the
second latch-up attempt, do not use the pump. Contact Smiths Medical Customer Service for further
assistance.
8. A brief message appears on the pump screen so you can verify the type of cassette you have attached.
9. To lock the cassette, insert the pump key into the cassette/keypad lock and turn
it clockwise into the locked position. “Cassette Locked” appears briefly in the
status bar.
Correct Incorrect
An uneven gap indicates
that the cassette is not
properly latched.