Operator's Manual
21
PumP SetuP
PumP SetuP
To install the batteries or rechargeable battery pack:
1. Make sure the pump is stopped or powered off. Using your fingers,
the pump key, or a coin, turn the knob on the battery door counter-
clockwise and open the battery door.
2. Hold the pump at an angle and
place 4 AA batteries in the pump,
from the bottom up (see picture).
Match the + and - markings on the
new batteries with the markings
on the pump.
– OR –
If using a rechargeable battery
pack, insert it into the pump as
shown.
3. Close the battery door and using your fingers, the pump key, or a coin, turn the knob on the battery door
clockwise to lock.
Note: If you put the batteries in backwards, the pump will not power up. Check the batteries, making
sure to match the + and - markings.
To install a CADD®-Solis Communication Module:
See the CADD Solis Communication Module Operator’s Manual to install the communication module into the
pump.
Replacing the Battery Door
If the battery door is removed or needs replacing (e.g., after removing the
CADD®-Solis Communication Module to use AA batteries instead), simply
snap the battery door onto the bar that is located on the pump (see picture).
– OR –