Example 3: Intermittent Boluses (IB) and PCA doses,
with Bolus Interval Type set as Bolus Interval
IB
PCA
dose
Time
Dosage
PCA
dose
denied
IB
IB
Next Bolus
delay
Intermittent
Bolus (IB)
Interval
IB
Delivery of PCA dose
moves out the delivery
of the next IB.
Bolus
Interval Type
(set as Bolus
Interval)
Intermittent
Bolus (IB)
Interval
Programming Flexibility
With the CADD®-Solis system, you have the flexibility to delay a PIB following a PCEA dose based on the
PIB interval or PCEA dose lock out interval.
Intermittent Bolus Examples
Dose (mL)
10
5
0
Bolus Interval : 1 hour
PCEA Dose Lockout : 15 min
Intermittent
Bolus
5:00
Intermittent
Bolus
6:00
PCEA
Dose
6:50
Intermittent
Bolus
7:05
7:00
Time
Dose (mL)
10
5
0
Bolus Interval : 1 hour
PCEA Dose Lockout : 15 min
Intermittent
Bolus
5:00
Intermittent
Bolus
6:00
PCEA
Dose
6:50
Intermittent
Bolus
7:50
7:00
Time
Bolus Interval Type Setting = PCEA Dose Lockout
– Intermittent boluses will deliver every hour
unless the time between a delivered PCEA Dose
and the next scheduled PIB is less than the PCEA
Dose Lockout time
Bolus Interval Type Setting = PIB Bolus Interval
– Intermittent boluses will deliver every hour
unless a PCEA Dose is delivered