Error Correction Maintenance
When errors occur, they should be corrected as shown below.
Display Contents Corrective action
Pressure being applied is
above the rating.
Memory data has probably
been corrupted in some way.
Operate within the rated
pressure range.
Perform auto zero adjustment.
Battery voltage is low. Replace the batteries.
Entire display flashes
Caution
Caution
Trimmer
Standard
pressure gauge
Compact
manometer
Compressor
Related Products Useful for Measuring Line Pressure
Valve type
Fluid
Proof pressure
Maximum operating pressure
Operating vacuum pressure
∗
Ambient and fluid temperature
Applicable tubing material
Note)
Option
2 port valve, 3 port valve
Air
1.5 MPa
1.0 MPa
–100 kPa
0 to 60°C
Nylon, Soft nylon, Polyurethane
Bracket
Finger valve
Series VHK
Specifications
Note) Use caution with soft nylon and polyurethane at the maximum operating
pressure. (For details, refer to Best Pneumatics Vol. 15.)
∗ For a vacuum application, use VHK2 (2 way valve).
JIS Symbol
Tubing material
Tubing O.D.
Self-seal fittings
Series KC
Applicable Tubing
Fluid
Maximum operating pressure
Proof pressure
Ambient and fluid temperature
Thread
Seal (Thread portion)
Adaptor for copper family
incompatible parts
Air
1.0 MPa
3.0 MPa
0 to 60°C
JIS B 0203 (Taper threads for piping)
JIS B 0211 Class 2 (Metric fine thread)
With seal (Standard)
Part C3604BD (Electroless nickel plated)
Specifications
Nylon, Soft nylon, Polyurethane
ø4, ø6, ø8, ø10, ø12
Mounting section
Nut section
Body
Stud
Chuck spring
Guide
Collet release bushing
Valve retainer
Stopper
Seal O-ring
C3604BD, PBT
C3604BD (Thread portion)
Stainless steel 304
C3604BD, Polyacetal (POM)
POM
POM
C3604BD, Polyacetal (POM)
NBR
Principal Parts Material
These products are convenient for measuring line pressure easily without the need to remove piping or stop supply pressure, etc.
2 port valve 3 port valve
A
II
P
RP
A
For details, refer to Best Pneumatics Vol. 15.
For details, refer to Best Pneumatics Vol. 15.
앬 Span calibration method
Do not touch the span calibration trimmer except when
performing span calibration.
1. Perform zero clear at atmospheric pressure.
2. Apply the maximum rated pressure, and calibrate the span while
comparing with a standard pressure gauge.
3. If the display value of the compact manometer is “0” after returning
to atmospheric pressure, then calibration is complete.
If the display value is not “0”, calibrate again by repeating steps 1
and 2.
앬 Replacing the batteries
When battery voltage becomes low the entire LCD will flash.
When the LCD flashes replace the batteries. Use 2 pcs. of type AA
dry batteries.
To replace the batteries, turn the power OFF and replace them
within approximately 30 seconds.
When not completed within 30 seconds, “Err” will be
displayed. In that case, perform zero clear once again.
In the event that the display runs out of control, remove the
batteries for one minute or longer, and then perform zero clear
again for inserting the batteries and turning on the power.
Switching between pressurization and atmospheric re-
lease can be easily performed by switching the control.
Pressure can be supplied or stopped by inserting or re-
moving a tube.
14-12-8
Series PPA