8 RESITIVE BRIDGE MODES (MODE 9 – 14)
8.1 Mode 9. Ub2: resistive bridge, ref. is V
bridge
, +/- 4% imbalance
In this mode, a resistive bridge can be measured where the ratio of the bridge output voltage
V
CD
and the
bridge supply voltage
V
AB
represents the physical signal.
The measurement range of the bridge imbalance is +/-4% in this mode.
The connection of the bridge to the UTI is shown in
Figure
13. The driving voltage across the bridge
V
EF
is
a square wave with amplitude
V
DD
. The frequency of this signal is 1/4 of the internal oscillator frequency.
Because force/sense wires are used, the bridge is measured in a 4-wire setup, as shown in Figure
13
(a).
The signals measured in the various phases are given in
Table 16
Figure 13. Connection of the resistive bridge to the UTI for the Ub2 mode in a 4-wire setup (a) and a 2-
wire setup (b)
During phase 2, the voltage across the bridge
V
AB
is measured. A very accurate on-chip voltage divider
divides this voltage by 32. This divider does not need calibration. After division,
V
AB
is processed in the
same way as
V
CD
.
Phase Measured voltages Output periods
1 V
offset
Toffset=N*K
2
*Voffset
2 V
ab
/32+V
offset
Tab= N*K
2
*(Vab/32+Voffset)
3 Vcd+V
offset
Tcd=N*K
2
*(Vcd+Voffset)
Table 16. Measured voltages of Ub2 mode