Users Manual Part 2
28
Guidance and MANUFACTURER’S declaration –
electromagnetic IMMUNITY
Guidance and manufacturer’s declaration – electromagnetic immunity
The Grid Pad 10s is intended for use in the electromagnetic environment specified below. The customer or the user of the Grid
Pad 15 should assure that it is used in such an environment.
Immunity test EC 60601 test level Compliance level Electromagnetic environment - guid-
ance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
6 Vrms 150 kHz to 80
MHz outside ISM bandsa
10 V/m
80 MHz to 2.7 GHz
3 Vrms
150 kHz to 80 MHz
6 Vrms 150 kHz to 80 MHz
outside ISM bandsa
10 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the Grid Pad 10s, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
80MHz to 800MHz
800MHz to 2.7GHz
where P is the maximum output power
rating of the transmitter in watts
(W) according to the transmitter
manufacturer and d is the recommended
separation distance in metres(m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey, a should be less than the
compliance level in each frequency
range b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.