11. EMC GUIDANCE
Guidance and manufacture’s declaration – electromagnetic emission
The No touch Infrared Body Thermometer is intended for use in the electromagnetic environment specified
below. The customer of the user of the No touch Infrared Body Thermometershould assure that it is used in
such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The No touch Infrared Body Thermometeruse RF
energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class B
The No touch Infrared Body Thermometer is
suitable for use in all establishments, other than
domestic and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacture’s declaration – electromagnetic immunity
The No touch Infrared Body Thermometeris intended for use in the electromagnetic environment specified below. The
customer or the user of No touch Infrared Body Thermometer should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Con:8 kV
Air:2,4,8,15 kV
Con:8 kV
Air:2,4,8,15 kV
Floors should be wood, concrete or
ceramic tile. If floor are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
Not applicable
Not applicable
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
Not applicable
Not applicable
Mains power quality should be that of
a typical commercial or hospital
environment.