Owner's manual

9
Leland Legacy Quick Guide
Terms »
Star button Ò
Scrolls through run time data and Setup options
Up and down arrow buttons ST
Toggle between display choices and increase or decrease sampling parameters in Setup
Button sequence
T Ò= press buttons individually
[ST] = press simultaneously
ÒSTÒ = security code, always press in sequence
Security code ÒSTÒ
Prevents unauthorized changes to the pump’s sampling program
Programming Sequences »
To activate pump (e.g., to change pump from Sleep to Hold):
Press any button.
To change pump from Hold to Run or Run to Hold:
Press [ST].
To reset accumulated data:
Press [ST], then ÒSTÒ. Press Ò until CLr displays then press [ST]; press Ò until End displays then
press [ST].
• To set pump ow rate:
Press [ST], then ÒSTÒ. Flow rate and SET ash. Press S or T to change ow rate. Press Ò until End
appears then press [ST] to save setting and place pump in Hold.
• To calibrate ow rate with standard calibrator:
Press [ST], then ÒSTÒ. Flow rate and SET ash. Press S or T to change ow rate. Press Ò once. ADJ
displays. Press S or T until desired ow rate is indicated on calibrator. When nished, press Ò until End displays
then press [ST] to save new setting and place pump in Hold. For CalChek Calibration, see operating instructions.
To change temperature scale from F to C or C to F:
Press [ST], then ÒSTÒ. Press Ò until temperature displays. Press S or T to switch units; press Ò until
End displays then press [ST] to save new setting.
To change atmospheric pressure scale (mm, mb, In):
Press [ST], then ÒSTÒ. Press Ò until pressure displays then press S or T to switch units; press Ò until
End displays then press [ST] to save new setting.
To change time scale (12 Hr/24 Hr/Dela):
Press [ST], then ÒSTÒ. Press Ò until 12 Hr, 24 Hr, or Dela displays then press S or T to switch units;
press Ò until End displays then press [ST] to save new setting. To set delayed start (Dela), see operating
instructions.
To change clock:
Press [ST], then ÒSTÒ. Press Ò until clock displays then press S or T to change ashing hour; press Ò to
move to minutes and S or T to change setting. Press Ò until End displays then press [ST] to save new setting.
To change the sampling time function:
Press [ST], then ÒSTÒ. Press Ò until ST L/min displays then press S to change ashing digit; press Ò
until
End displays then press [ST] to save new setting. To delete, follow above steps and press T until 0 appears.
Exit Setup.
Note: When in Setup, choosing Esc instead of End will exit Setup without saving new settings.
SKC Inc., 863 Valley View Road, Eighty Four, PA 15330 • www.skcinc.com