User guide

8
SKC Real Time Monitor Window
Flow Calibrate
The Flow Calibrate bu ons (Figure 6) allow calibration of the ow rate displayed on
the pump LCD to the ow rate displayed on a primary standard calibrator.
The Approx Correction (approximate correction) is the di erence between the ow
rate displayed on the pump and the ow rate displayed on the calibrator. When
a pump is connected to a calibrator the ow rate is determined by the calibrator
display and not the pump display.
Adjusting the Flow Correction
To adjust the approximate correction rate, click and hold the sliding bu on, then
move the bu on to the le (decreases correction) to adjust ow when the calibrator
display is higher than the pump ow rate display or to the right (increases correction)
to adjust ow when the calibrator display is lower than the pump ow rate display.
The range of correction is ± 20 ml/min. An alternate method is to click the le and
right arrow bu ons to move the sliding bu on. The value changes as the bu on
slides le or right.
The pump should be calibrated before each sample run. Once calibrated, all volume displays will be accurate for that
ow rate. DataTrac will always display the last approximate correction value to which the connected pump has been
set. Changing the ow se ing on the pump clears the correction value (sets it to 0.0).
When calibrating, the ow rate displayed on the calibrator changes; the ow rate displayed on the
pump LCD does not change.
Reset Button
To reset the correction value to 0.0 mL/min, click once on the Reset bu on.
Multiple Pumps Checkbox
Turn off Comm Checking before setting up multiple pumps.
See Comm Checking Buttons on page 15.
If you wish to perform ow calibration, reset volume temperature and time,
or change parameter display scales for multiple pumps, click the Multiple
Pumps checkbox in the Flow Calibrate section of the Real Time Monitor
window.
Once checked, a separate window will open (Figure 6A). Set parameters as
desired for one pump, disconnect the interface cable from the pump, and
insert the cable plug into the interface port on another Pocket Pump. Once
all pumps are set up, click the Multiple Pumps checkbox to deselect this
option. You will be returned to the Real Time Monitor window.
Using Flow Calibrate
Set the pump to the desired ow rate.
Connect the inlet port of the pump to a primary standard calibrator and read the ow on the calibrator display.
Set the approximate correction as needed by clicking and holding the sliding bu on and moving it le to decrease
correction or right to increase correction and recalibrate. The ow rate that is displayed on the calibrator should change
as result of this operation.
Example: The desired ow rate of the pump is 150 mL/min. Set the pump to 150 mL/min. If the calibrator reading is
157 mL/min, move the sliding bu on to the le to adjust the approximate correction to -7 mL/min and recalibrate. Repeat as
necessary until the ow rate displayed on the calibrator is within the required tolerance.
Figure 6. Flow Calibrate
Click here to increase
the pump ow rate.
Click here to decrease
the pump ow rate.
Figure 6A. Multiple Pumps Checkbox