User guide

Always replace the used foam disk in the back cap with a clean disk a er cleaning the housing.
This sampler only samples elemental mercury (Hg) in the vapor phase and does not sample
organic mercury compounds or particulate ma er.
Take precautions when sampling for mercuric chloride. It can collect on the face of the badge and
interfere with sampling.
Interferences During Sampling
Ambient Temperature
The sampling rate will vary with ambient temperature changes which a ect the
di usion co-e cient. This e ect is small but may be signi cant if sampling at un-
usually high or low temperatures. The di usion coe cient (D) is a function of the
absolute temperature (T) by the relation D T
1.5
. This factor is taken into account
in the calculation used in the OSHA method (see page 4).
Wind Velocity
The sampling rate will remain substantially constant over a range of wind velocity
from 25 to 750 /min. In very still air conditions (below 25 /min), the sampling
rate will drop by up to 30%. If very high wind velocities (in excess of 750 /min)
are expected, passive samplers should not be used.
Ambient Pressure Changes
The sampling rate varies with ambient pressure changes which a ect the di usion
co-e cient. The di usion coe cient is inversely proportional to the ambient pres-
sure. This factor is taken into account in the calculation used in the OSHA method
(see page 4).
Presence of Other Gases
The sampling of mercury on the sorbent has been shown to be una ected by the
presence of other gases, including chlorine. Note, however, that complex interac-
tions in the atmosphere may reduce the total free elemental mercury present to be
sampled.
Analysis
Consult OSHA Method ID 140 and MDHS 16 for detailed analytical procedures.
Preparation of Samples
Open the sorbent capsule and carefully pour the sorbent into a 50-ml volumetric
ask. Add 5 ml of concentrated nitric acid followed by 5 ml of concentrated hydro-
chloric acid. Allow to stand for approximately 30 minutes or until the sorbent is
fully dissolved. Stir gently if necessary. Add approximately 15 ml of distilled water
to bring total liquid quantity up to 25 ml.
When Anasorb C300 is dissolved, a milky suspension results. This is an insoluble
residue and must be removed by ltration or centrifugation before analysis.
Take aliquots from this solution and analyze at 253.7 nm by cold vapor atomic
absorption.
Prepare several “blanks” by dissolving fresh, unexposed capsules of the same lot
number as the test specimen. These “blank” values represent the background level
of mercury in the sorbent and should be subtracted from all readings.
Mercury Desorption
The basic mechanism of desorption consists of adding an excess of stannous chloride
to the aliquot of sample solution. This causes the mercury to be released as elemental
mercury which is swept into the analyzer by an air or nitrogen stream.
oc