Instruction Manual

Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
IOM Personal Samplers and
IOM Samplers with MultiDust
Description
The patented* IOM Sampler is a sampling head that houses a reusable two-part
lter casse e with speci ed 25-mm lter for the collection of inhalable airborne
particles. The lter and casse e assembly are pre and post-weighed as a single unit
for gravimetric analysis. The IOM Sampler and casse e are available in conductive
plastic or stainless steel. The stainless steel model is ideal for sampling vapor-phase
isocyanates followed by chemical analysis. The IOM Sampler meets the following
U.S. and international standards:
ACGIH
®
sampling criteria for inhalable particulate mass
NIOSH Method 5700 for formaldehyde on dust
Australian standard for inhalable particulate
Preferred sampler for HSE Method MDHS 14/3 for inhalable and respirable dust
Complies with MDHS 25/3 for organic isocyanates (with stainless steel model)
Complies with MDHS 6 for lead (with accessory single-hole head)
• ISO
®
/CEN health-related fractions of bioaerosols
OSHA equivalent method for particulates not otherwise regulated (PNOR)
The sampler is supplied with a casse e cover and casse e transport clip to prevent
contamination during transport.
* U.S. Patent No. 4,675,034
Reference: OSHA le er November 8, 2011
IOM with MultiDust Foam Disc
The IOM can be used to sample for inhalable and respirable fractions simultane-
ously by using a lter and a MultiDust Foam Disc (polyurethane foam). In addition,
a polycarbonate lter can be combined with a MultiDust Foam Disc for e ective
size-selective bioaerosol sampling and be er microorganism survivability when
compared to lter-only sampling.
1
Conductive plastic IOM
Stainless steel IOM