User guide

3
Impactor Preparation
Cleaning
For optimum performance, the IMPACT Sampler inlet, exhaust, and lter casse e
should be cleaned a er ve runs or upon a noticeable buildup of material. This
will remove oil buildup from the top of the lter holder and other residue built
up from frequent sampling. Disassemble the impactor and wash parts in water
with a liquid detergent or soap. Rinse and air-dry all parts thoroughly in a clean
environment.
Caution: Do not place any mechanical object in the inlet nozzles.
O-ring Care
Visually inspect the condition of the BUNA-N exhaust O-ring (see illustration on
page 2 for location). Ensure the O-ring surface is smooth (i.e., without cracks, cuts,
or other damage). Ensure the O-ring is ed properly in its channel. Replace the
exhaust O-ring if there is apparent damage, stretching, or thinning. It is recom-
mended that the PTFE inlet O-ring (2 in PM Coarse model) be replaced by the
manufacturer only.
Inserting a Collection Filter into the IMPACT Sampler
The IMPACT Sampler will arrive already assembled. Disassemble it to insert
collection lter.
Unscrew inlet from exhaust. Remove lter casse e.
3a. Use the lter casse e opener
accessory to open lter casse e.
3b. Slide the lter casse e horizontally
into the "U" of the opener until the
two halves of the casse e loosen.
Gently pull halves apart.
Ensure the stainless steel support
screen is in place in bo om of
lter casse e.
Using forceps accessory, place a
preweighed 47-mm lter on the
support screen.
Press lter casse e top into
lter casse e bo om.