Manual

Cleaning
Using mild soapy water, clean all parts of the Bu on Sampler before further sam-
pling. The Bu on Sampler can be brushed and/or wiped with a clean dust-free
tissue, air dried, blown dry, or wiped with isopropyl alcohol. Take particular care
in cleaning the PTFE O-ring.
The Bu on Sampler may also be autoclaved up to approximately 273.2 F (134 C).
Prepare the Bu on Sampler before autoclaving by removing the collar clip from
the sampler using a Phillips head screwdriver. Also remove the BUNA-N O-ring
from the sampler body (see page 2 for exploded view of sampler).
References
Kalatoor, S., Grinshpun, S., Willeke, K., “New Aerosol Sampler with Low Wind Sensitivity and Good Filter
Collection Uniformity,” Atmospheric Environment, Vol. 29, No. 10 (1995), pp. 1105-1112
Hauck, B., Grinshpun, S., Reponen, A., Reponen, T., Willeke, K., Bornschein, R., “Field Testing of New
Aerosol Sampling Method With a Porous Curved Surface as Inlet,” American Industrial Hygiene Association
Journal, Vol. 58, No. 10 (1997), pp. 713-719
Aizenberg, V., England, A., Grinshpun, S., Willeke, K., Carlton, G., “Metal Exposures Among Abrasive
Blasting Workers at Four U.S. Air Force Facilities,” Applied Occupational and Environmental Hygiene,
Vol. 15, No. 10 (2000), pp. 766-772
Aizenberg, V., Grinshpun, S., Willeke, K., Smith, J., Baron, P.A., “Performance Characteristics of the Bu on
Personal Inhalable Aerosol Sampler,” AIHA Journal, Vol. 61, May/June 2000, pp. 398-404
Optional Accessories
Description Cat. No.
Button Sampler, requires a 25-mm lter 225-360
Button Sampler Pump Kit includes Button Sampler, standard
XR5000 Sample Pump, single charger with cable, 3 feet of
Tygon tubing, and calibration adapter, requires a 25-mm lter
100-240 V
210-4121
Abrasive Blasting Kit includes Button Sampler and protective
shield, requires a 25-mm lter 225-367
Button Sampler Calibration Adapter 225-361
Protective Shield, for abrasive blasting environments 225-366
Filter Transport Case, for 25-mm lters, conductive plastic 225-67
Replacement O-ring, PTFE P32265
Filters (25 mm)
Cat. No.
Glass ber, 1.0 μm, pk/500 225-702
PTFE with PMP support (Te o
®
),
**
3.0 μm, pk/50 225-1711
Polyvinyl chloride (PVC), 5.0 μm, pk/100 225-8-04
Mixed cellulose ester (MCE), 1.2 μm, pk/100 225-1912
Gelatin,
#
sterilized, pk/50 225-9551
A pore size of 1.0 μm or larger is recommended for use with the Bu on Sampler due to back pressure
limitations of personal samplers.
Back pressure on PTFE lters can vary within the same lot.
# Gelatin lters dissolve when placed on agar.
** Maximum operating temperature for the Cat. No. 225-1711 PTFE lter with PMP support is
464 F (240 C) based on the PMP support ring.
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC’s sole liability and the buyers exclusive remedy. To view the com-
plete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/
warranty.asp.
Form 3780 Rev 1303