Manual

4. For area sampling, position the Bu on Sampler to avoid cross-dra s and
direct projection of the particles into the inlet.
5. At the end of the sampling period, stop the pump and record the stop time.
Sample Shipment
1. When sampling is complete, very gently unscrew the sampler inlet and use
forceps to remove the PTFE O-ring.
2. Using forceps, carefully remove the lter from the sampler.
3. Package for shipment as directed by the method or laboratory.
4. Ship samples with pertinent sampling information to an accredited laboratory
for analysis.
Analysis
Dependent upon the hazard of interest, samples taken with the Bu on Sampler
can be analyzed using gravimetric analysis, analytical techniques such as x-ray
uorescence, atomic absorption (AA), or inductively coupled plasma (ICP).
Growth culture, microscopy, and other assays may be used for the determination
of fungal spores, endotoxin, and bacteria.
If gravimetric analysis is used, weigh the lter and subtract the preweight from the
postweight. The net weight will be the dust collected.
Using the Button Sampler as an Abrasive Blasting
Sampler for Heavy Metals
SKC has designed a protective shield for the Bu on Sampler for use when sampling
in abrasive blasting environments. The shield protects the lter from shredding
or being overloaded by large particles thrust onto the sampler during
blasting. Tested by the U.S. Military during recent abrasive blasting operations, the
Bu on Sampler with protective shield (Abrasive Blasting Sampler for Heavy
Metals) withstood very high particle concentrations and provided recoverable
samples and meaningful exposure information. The shield is very easy to use,
simply push it onto the sampler inlet and sample (see Sampling).
‡ The Bu on Sampler with shield is not suitable for sampling silica. While silica is commonly found in
abrasive blasting environments, sampling for silica using OSHA Method ID 142 requires a cyclone.
Using the Button Sampler to collect Inhalable
Bioaerosols
The patented SKC Bu on Aerosol Sampler provides superior collection of
inhalable particles including bacteria and fungal spores. Use the Bu on Sampler
with a 25-mm membrane lter such as MCE or PVC to collect bioaerosols for
viable or non-viable analysis. Using SKC gelatin lters with the Bu on Sampler
increases the survival of stress-sensitive microorganisms during sampling. The
Bu on Sampler is used with a sample pump capable of 4 L/min for personal
inhalable particulate sampling following the ACGIH/ISO criteria.
Note: Gelatin lters dissolve when placed on agar.