Manual

15
Turn the ow adjust screw (Figure 4, #2) for the appropriate
port of the low ow holder until the desired ow rate is
achieved through the tube. Turn clockwise to decrease the
ow.
Do not adjust the ow on the pump. Adjust the ow only by
using the ow adjust screw on the low ow holder.
Do not exceed 500 ml/min ow rate per tube.
When the desired ow is set on the
initial tube, place pump in Hold by
pressing Flow and Battery Check.
Remove the calibrator from the tube
and connect to the exposed end of
the next sorbent tube. Press Flow and
Battery Check and repeat the ow
adjustment process until all tubes are
ow calibrated. Changing the ow on
one tube will not affect the ow rate
through the remaining tubes.
Do not exceed 500 ml/min ow rate per tube.
For tri and quad models, rst rotate each anti-tamper cover
(Figures 4 [on page 13] and 5) to expose the ow adjust
screws, then adjust the appropriate screw until calibrator
indicates the desired ow.
When the ow rate is set for each tube, press Flow and
Battery Check to place the pump in Hold and disconnect the
calibrator.
Replace the sampling media used for calibration with
unexposed media for sample collection. Use a protective tube
cover to prevent tube breakage.
If sampling with fewer tubes than number of ports, insert
unopened sorbent tubes in the empty ports to seal them.
Screw
Screw
Figure 5 -
Cut-away of Tri/Quad Low
Flow Holder
2
cont’d