Setting or Verifying Flow Rate
When performing multiple-tube sampling using an adjustable
low ow holder (dual, tri, or quad), ensure the regulator has been
activated and the pump ow rate is set at 1.5 L/min. The maximum
ow rate through any one tube is 500 ml/min*. Calculate the sum
of all tube ow rates. If the sum is 1000 ml/min, proceed with
calibration and sampling without any further adjustment to the
pump ow rate. If the sum is >1000 ml/min, set the pump ow rate
15% higher than the sum of tube ow rates.
* Back pressure across some sample tubes can be higher than average. In
these instances, the maximum ow rate of 500 ml/min per tube may not be
achieved.
Before use, allow pump to equilibrate after moving it from one
temperature extreme to another.
Ensure pump has run for five minutes before proceeding with
calibration.
Ensure the pump is set for low ow. See Activating the Regulator,
page 9.
Connect the adjustable low flow holder (Figure 4, page 13)
to the pump intake (Figure 2, #9) using 1/4-inch Tygon tubing.
Insert an opened sorbent tube into each rubber sleeve of the low
ow holder (Figure 4, #3 and 4) with the arrow on the tube pointing
toward the holder.
If sampling with fewer tubes than number of ports, insert
unopened sorbent tubes in the empty ports to seal them.
Note the ow rates speci ed by each sampling method and
add them together. If the sum is 1000 ml/min, proceed to
the next step. If the sum is >1000 ml/min, multiply the total
tube ow rate by 1.15 and set the pump for that ow rate.
Connect a calibrator to the exposed end of a sorbent tube,
loosen the brass ow adjust screw on the low ow holder, and
activate the pump by pressing Flow and Battery Check.
Connect holder to
pump intake and tube
inlet to calibrator.
continued on page 15