Manual

11
Setting or Verifying Flow Rate
For a diagram of the pump, see Figure 1, page 5.
Before use, allow pump to equilibrate after moving it from one
temperature extreme to another.
Ensure pump has run for ve minutes before proceeding with
calibration.
Connect a single adjustable low ow holder (Figure 3) to the pump
intake (Figure 2, #9) using 1/4-inch Tygon tubing.
Insert an opened sorbent tube (Figure 3, #3) into the rubber sleeve
(Figure 3, #2) of the low ow holder with the arrow on the tube
pointing toward the holder.
Connect a calibrator to the exposed end of the sorbent tube.
Loosen the brass ow adjust screw on
the low ow holder. Activate the pump
by pressing Flow and Battery Check.
Adjust the ow rate by turning the
ow adjust screw (Figure 3, #1) on the
holder until the calibrator indicates the
desired ow.
Do not adjust the ow on the
pump. Adjust the ow only by
using the ow adjust screw on
the low ow holder.
When the desired ow is set, place
pump in Hold by pressing Flow and
Battery Check. Disconnect the calibrator.
Figure 3
Single Adjustable
Low Flow Holder
with sample tube
1 Flow adjust screw
2 Rubber sleeve
3 Sorbent tube
Air ow
Calibration train with
tube in low ow holder
Flow adjust
screw
Turn screw on
low ow holder
to adjust ow.
continued on page 12
3