Manual

10
Activating the Regulator
Remove the tamper-resistant cover.
Start the pump using the on/off switch
(Figure 2, #4). Press Start/Hold (Figure
2, #3). Press Flow and Battery Check
(Figure 2, #2). Adjust ow using the
ow adjustment screw (Figure 2,
#7) until the built-in rotameter reads
1.5 L/min. The LCD should indicate
BATT OK in the upper left corner (if
it doesn’t, recharge the battery). Press
Flow and Battery Check to place the
pump in Hold.
Remove the cap screw covering the regulator valve (Figure 2, #14)
and turn the exposed screw four to ve turns
counterclockwise.
Replace the cap screw. The pump is now set for low ow.
For low ow,
turn valve screw
counterclockwise.
2