PowerFlex Charging System Cat. No.
PowerFlex 5-station Charger
includes power supply and cord.
Purchase pump-speci c cables separately; choose from list below
100-240 V 223-1000
PowerFlex Single Charger
includes power supply. Purchase pump-
speci c cables separately; choose from list below. 100-240 V 223-2000
PowerFlex Pump-speci c Cables
Cat. No.
AirChek
®
2000 223-1001
Universal XR (5 cell) 223-1002
Universal XR (4 cell, MSHA) 223-1003
AirChek 52 223-1004
222 223-1005
Accessory Cat. No.
Nylon Carry Case, for 5-station PowerFlex, includes
padded shoulder strap and compartment for cables
224-903
PowerFlex Technology
Charging NiCad/NiMH batteries is generally accomplished by applying constant current to the battery.
SKC PowerFlex Chargers use a technology that is new to industrial hygiene equipment batteries.
Re ex charging
delivers pulses of current instead of constant current to charge batteries. In between
pulses, PowerFlex applies short reverse pulses that discharge the battery slightly and short pauses.
This eases the battery into a full and uniform charge. It helps to prevent dramatic increases in internal
temperature and pressure that can damage the battery and shorten its life.
Patented by Galaxy Power, Inc.
Note: Discharging the pump battery before charging is not necessary with this system
because the charging technology does not promote memory. However, if the user
desires to discharge before charging, simply run the pump until it stops, then
charge the battery.
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole
liability and the buyers exclusive remedy. To view the complete SKC Limited Warranty and Return
Policy, go to h p://www.skcinc.com/warranty.asp.
Troubleshooting
If the LED does not light when a ba ery pack is plugged into a port or the LED
does not turn o a er a ba ery pack is unplugged, follow the steps in the table.
Step Action No result, Action Yes result, Action
1 Keep the unit powered on. If the battery is deeply
discharged, the LED may take up to 30 minutes
to light or turn off.
Go to Step 2 Proceed with operation
2 Check fan. Is it running?
a. Check power to unit and/or
b. Move unit to a different wall outlet
Go to step 3 Proceed with operation
3 Move battery to different port Go to Step 4 Proceed with operation
4 Unplug charger from its power source for 20
seconds. Plug it back into power source.
Go to Step 5 Proceed with operation
5 Replace pump-speci c cable Go to Step 6 Proceed with operation
6 Try a different battery pack Call SKC Proceed with operation —
previous battery pack may
be damaged; call SKC.