Instruction Manual

8
User Setup Functions
Function Overview
User setup functions are listed below in the order in which they display. Note that the
CLr function for clearing accumulated run time is only available when accumulated
run time exists.
No Accumulated Run Time Accumulated Run Time
CLr
ADJ Flow ADJ Flow
Set Timed Run Set Timed Run
Set Run Delay Set Run Delay
Clearing Accumulated Run Time
1. With the pump in a non-running state (no ashing blue
LED), press .
2. Press [] at CLr display to clear accumulated run
time. Pump is ready to run.
CLr will not cancel Timed Run or Run Delay time settings (see Canceling a
Timed Run and/or Run Delay).
Changing the timed run and/or run delay settings in user setup functions will
automatically clear accumulated run time.
Changing the ow rate in user setup functions will not clear accumulated
run time.
Setting Flow Rate
1. With pump in a non-running state (no ashing blue
LED), press .
2. Connect pump inlet to a calibrator.
3. Press until ADJ and Flow ash on display.
4. Press to increase or to decrease ow. Dashed
lines will move up and down on the LCD to indicate
direction of adjustment. Flow rate will not display
on the pump LCD. Observe the calibrator for ow
reading.
5. Press [] to accept ow setting and to exit user setup functions.
See Calibration for instructions on calibrating pump ow rate.
Setup