Instruction Manual

7
Calibration NIOSH 0600 with ARD
Accuracy + 10%
Precision 0.02 mg/m
3
Sensing range 0.01 to 200 mg/m
3
Particle size ranges 0.1 to 10 micron Respirable (Use IOM and selective foam)
0.1 to 50 micron Thoracic (Use IOM and selective foam)
0.1 to 100 micron Inhalable (Use IOM inlet)
Recording time 1 second, 1 minute and 10 minute averages
Memory 21,500 data points
Locations Up to 999 storage locations
Data display 20 character, 4 line backlit LCD
Output RS-232
Operating temperature 0°-50°C
Humidity range 95% non-condensing
Battery NiMH rechargeable battery
Battery life 8 hours minimum when fully charged
Size 18 x 8.. x 4.5 cm protruding sockets and belt clip excluded from measurements
Weight 780 grammes
Software Windows
TM
98, 2000, NT, XP
Flow rate 2.0 litres per minute using external pump (not included)
Optional Sampling Pump Sidekick Pump. Part #224-50 or 224-52TX
ON/OFF button:
Turns the SPLIT2 on and off
ENTER button: Activates the SELECTED
option. Shown by an arrow to the left of the
menu line
Scrolls the Selection Arrow up one line
Scrolls the Selection Arrow down one line
SPECIFICATION
CONTROLS & LCD
Run
Review Data
Special Functions
Auto Zero
Example of 4 line LCD display of the SPLIT2
ON/OFF button ENTER button
UP button
DOWN button
Switching on the SPLIT2
SPLIT2 version 3.2 onwards features a manual on/off switch on
the side, located adjacent to the charger socket.
This must be on for the control panel to become activated.
We recommend switching off by this method, to deactivate the
keypad at the end of sampling or whenever the unit is put away.
The manual switch can also be used as a reset if required.