www.skcltd.
Purchase Details and Service History Thank you for choosing an SKC product. Your purchase is covered by our warranty, details of which can be found inside the rear cover of this manual. Product Model Number Product Serial Number Date of Purchase SKC recommends annual servicing of this product. The first service is due one year from the date of purchase, and then at yearly intervals on this date.
Contents Specifications..........................................................................................................................................................2 General Information 1) Pump Models..............................................................................................5 2) Care of the Flite 2 Pump..............................................................................5 3) The WEEE Directive...........................................................................
Specifications Dimensions.....................................................178mm W x 115mm H x 206mm D Weight............................................................2.24 kg (without battery) 3.56 kg (fitted with P901201 3.2Ah battery) 4.84 kg (fitted with P901202 7Ah battery) Casing IP rating..............................................IP20 Flow range.....................................................2 - 26 litre/min Maximum sampler back pressure...................
Specifications Pump Features Battery capacity options Choose from two capacities of internal battery - 3.2Ah (Part No. P901201), for applications where a lightweight pump is desirable, or 7Ah (Part No. P901202), where longer run times or multiple samples on a single charge are required. Intelligent battery charging The SKC lead/acid battery charger (Part No. 901-210), provides optimum charging in the minimum possible time.
Specifications Typical Pump Run Times Filter Type 25mm 0.8µm MCE 25mm 1.2µm MCE 25mm GFA Flow Rate Run Time P901201 3.
General Information 1) Pump Models 901-2011 Standard single port pump 901-2012 Twin port pump for through flow applications 2) Care of the Flite 2 Pump • • • • The Flite 2 pump is supplied complete with rubber dust caps for the inlet/outlet hosetails and a screw-in dust cap for the charging/power socket. Please ensure that these are fitted at all times when the pump is not in use or being charged.
The Flite 2 Pump Inlet hosetail LCD screen Charging / power socket Keypad Casing screws Outlet hosetail (901-2012 model only) Carry handle Page 6 Screw-in cap 901-FTM Issue H Casing feet www.skcltd.
Getting Started 1) Fitting the Battery The Flite 2 pump is supplied without a battery. Refer to the instructions supplied with each battery for details of how to fit it into the pump casing. Note - The toolkit (Part No. P901203) supplied with the pump is required to fit the battery. 2) Charging the Battery Prior to first use the battery should be fully charged, ideally overnight. The Flite 2 pump must only be charged using the correct SKC charger (Part No. 901-210).
Getting Started 3) Use of the Keypad and LCD Screen Key functions - Switches between menu options and selects digits to adjust when entering times. - Enters the selected menu option or run time. and - Used to enter the sample run times and set the pump flow rate. The available options for each menu item are displayed on the bottom line of the LCD screen. The currently selected option is indicated in capitals, with the remaining options in lower case as shown. Press the as shown.
Getting Started On number entry menu screens such as the sample run time screen, each digit of the hours and minutes is entered individually. The currently selected digit is identified by being underlined. Press the key to scroll through the digits as shown. Use the and keys to enter the required value for the selected digit as shown. Press the key to accept the run time. The next menu screen will be shown on the LCD screen as shown. www.skcltd.
Programming the Flite 2 Pump 1) Waking the Pump from Sleep Mode When not in use the pump should be put into ‘Sleep’ mode to conserve battery power. The LCD display appears as shown when in Sleep mode. To ‘wake’ the pump from Sleep mode press the keypad keys in the sequence , , . The pump will display the ‘SET UP NOW ?’ screen.
Programming the Flite 2 Pump Select the ‘YES’ option in the ‘SET UP NOW ?’ menu screen, and press to display the next menu screen. SET UP NOW ? YES/no Æ SET FLOW NOW ? YES/no Æ SET FLOW USING >>> Press to increase the pump flow rate. More arrow symbols are displayed on the LCD screen to indicate the increased flow rate. Æ SET FLOW USING >>>>>>>>>>>> Press to decrease the pump flow rate. Less arrow symbols are displayed on the LCD screen to indicate the reduced flow rate.
Programming the Flite 2 Pump Having set the required pump flow rate disconnect the flow calibrator from the sampling head inlet. Remove the filter from the sampling head and replace it with a fresh new filter. Fit the sealing cap to the sampling head inlet to prevent contamination of the filter prior to sampling. 3) Manual Operation Having set the pump flow rate to the desired level, the ‘RUN PARAMETER’ menu screen is displayed.
Programming the Flite 2 Pump To enter a sample run time select the ‘YES’ option, and press to display the next menu screen. Alternatively, select the ‘NO’ option, and press to return to the ‘RUN PARAMETER’ menu screen. ENTER RUN TIME ? YES/no Æ / Use the , and keys to enter the required sample run time in hours and minutes. Press to save the sample run time and display the next menu screen. To enter a start delay time select the ‘YES’ option, and press to display the next menu screen.
Programming the Flite 2 Pump Press to save the start delay time and display the next menu screen. Æ REPEAT RUN AND D YES/no To program repeat sample run and delay times select the ‘YES’ option, and press to display the next menu screen. Alternatively, select the ‘NO’ option, and press to display the ‘REVIEW DATA ?’ menu screen. Refer to section - Operation 5) Reviewing Sampling Programs, for further details.
Programming the Flite 2 Pump To check the pump flow rate select the ‘YES’ menu option and press . The pump will run for 1 minute allowing the flow rate to be checked with a flow calibrator. Alternatively, select the ‘NO’ menu option and press to display the ‘TIMER MODE’ screen. Æ When the flow rate has been checked press the next screen. to display Æ If the flow rate was correct select the ‘YES’ menu option and press to display the ‘TIMER MODE’ screen.
Programming the Flite 2 Pump If the sample is to be carried out at a later time, put the pump into Sleep mode by selecting the ‘NO’ option and pressing . When ‘MANUAL’ mode is selected, the ‘START NOW ?’ menu screen includes the menu option ‘ESCAPE’. Selecting the ‘ESCAPE’ menu option and pressing displays the ‘RUN PARAMETER’ menu screen. Page 16 901-FTM Issue H START NOW ? yes/NO/escape Æ SAMPLING PUMP SLEEPING START NOW ? yes/no/ESCAPE Æ RUN PARAMETER MANUAL/time/exit www.skcltd.
Operating the Flite 2 Pump 1) Manual Run Having programmed the desired flow rate and fitted a new filter into the sampling head, place the pump at the sampling location, and ensure that the sealing cap is removed from the sampling head inlet. If the pump is in Sleep mode, wake it and either the USE LAST SETUP ? ‘REVIEW DATA ?’, or if the pump program has already been YES/no used to take a sample, the ‘USE LAST SETUP ?’ screen is displayed.
Operating the Flite 2 Pump + To stop the pump at the end of the required sample period, press the and select keys simultaneously. The LCD screen will appear as shown. Press the keys in the sequence , ‘CHECK FLOW ?’ screen is displayed. , and the Æ ELAPSED TIME 001:00:03 STOP Æ Æ Æ CHECK FLOW ? YES/no If required the flow rate can be checked to verify that it has remained within the required tolerance during the sample run.
Operating the Flite 2 Pump 3) Timed Run with Delayed Start Having previously programmed a timed run with delayed start, start the pump as detailed for manual run. On accepting the ‘YES’ option in the ‘START NOW ?’ screen the ‘TIME TO START’ screen is displayed. This screen displays the start delay time in hours, minutes and seconds, and counts down to ‘000:00:00’. TIME TO START 000:29:52 At the end of the start delay time the pump will automatically start and the ‘ELAPSED TIME’ screen is displayed.
Operating the Flite 2 Pump At the end of the first start delay time the pump will automatically start and the ‘ELAPSED TIME’ screen is displayed. Again the screen also displays the sample repeat number and the total number of repeats. At the end of the first run period the pump will automatically stop and the ‘TIME TO START’ screen is again displayed. Note that the sample repeat number has incremented.
Operating the Flite 2 Pump To restart the pump from Hold, press the keys in the sequence , , . The pump will restart and the elapsed time will continue to increment. Cancelling a Run - In any operating mode the run can be cancelled at any by point pressing the and keys simultaneously. The pump will stop and the elapsed run time is retained on the LCD screen. www.skcltd.
Troubleshooting Should the Flite 2 pump detect that the battery is at a low charge state a low battery message will be displayed on the LCD screen. Charge the battery fully using the correct SKC charger before use. The Flite 2 pump will display a general fault message on the LCD screen in the event of any error, as shown. Possible Fault BATTERY TOO LOW OR ELAPSED TIME 003:24:17LOW BAT ELAPSED TIME 000:16:53 FAULT Corrective Action 1. Check media for heavy loading or damage. Sample media fault 2.
Battery Care The Flite 2 pump utilises sealed, valve regulated, lead/acid batteries. This type of battery requires no maintenance and when used in accordance with manufacturer’s instructions should give a life of approximately 400 charge/discharge cycles (based on 50% depth of discharge). Battery life is affected by a number of factors, but the most significant of these are overcharging and the depth of discharge in use.
Flite 2 Pump Components Casing foot 2 1 Casing screw Casing lid Casing screw 2 Pump head & motor assembly Battery clamping bracket, nuts & washers Bulkhead connector & back nut 5 Keypad 6 Inlet/outlet hosetail 3 Battery Sealing cap Charging socket cap Power/charging socket connector Page 24 Casing base Carry handle Casing foot 901-FTM Issue H Casing screw 2 Main control PCB assembly 4 1 www.skcltd.
www.skcltd.
Flite 2 Pump Replacement Parts Item Part No. Description Item Part No. Description 1 P901111 Casing feet (pack of 4) 11 P901116 Valve housing (pack of 4) 2 P901110 Casing screws (pack of 10) 12 P901208 Pump head inlet tube (pack of 2) 3 P901201 Battery 12V 3.
Flite 2 Pump Accessories Part No.
Warranty Information Limited One Year Warranty 1. SKC warrants that this instrument, and each of its component parts, provided for occupational health and safety and environmental applications is free from defects in workmanship and materials under normal use for a period of one (1) year.
Warranty Information to the buyer. The remedy will consist of service or replacement of the instrument, or refund of the purchase price, at the option of SKC. However, SKC will not elect refund unless it is unable to provide replacement and repair is not commercially practicable. 5. The terms of this warranty begin on the date the instrument is delivered to the Buyer and continue for a period of one (1) year.
901-FTM Issue H September 2012 A member of the SKC global group of companies SKC Limited 11 Sunrise Park, Higher Shaftesbury Road, Blandford Forum, Dorset DT11 8ST United Kingdom T: +44 (0) 1258 480188 F: +44 (0) 1258 480184 E: info@skcltd.com W: www.skcltd.