Installation manual
Simrad SP90
140
851-164510 / Rev.E
10.3 Speed log
Introduction
The speed log can come from three different sensor types. Tick
off for the type which will be connected.
• Pulse log (200 pulses/nm)
• Speed log (RS-232 serial line)
• (D)GPS
Refer to the selected speed log source in the following text.
Pulse log (200 pulses per nautical mile)
The pulse log input shall be connected to terminal TB8 inside
the Sonar Interface Unit.
Observe the following procedure to test the pulse log input:
1 Access the I/O Setup menu.
2 Select Sensors in the I/O Setup menu, then Speed and
finally Speed Log.
3 Check that the settings in the Sensor Config menu
correspond to the sensor settings table. Note that the pulse
log must use serial port no.9.
4 Press Close to exit the Sensor Config menu.
5 Check that the speed readout in the Status menu
corresponds to the vessel’s speed log.
The pulse log connection and interface circuit is located in the
Sonar Interface Unit.
An oscilloscope can be connected to the test point TP33 to
check if the pulse log signal is present through the comparator
IC30 (TP34 is GND). If not, try to adjust the potentiometer
R64. In case of noise problems, R64 can be adjusted for noise
limitation.
Related topics
→ Location of TB8, page 88.
→ Access to the I/O Setup, page 137.
→ Sensor settings table, page 137.
→ Pulse log interface, page 89.