Installation manual

Operationalprocedures
6ClicktheSetupiconundertheMainmenutoopentheSetupmenu.
Setupmenuonpage112
7ClickCalibration.
8IntheCalibrationdialog,clickStartCalibration.
9WaitforthePI50systemtodothecalibration.
WhentheCalibrationdialogcloses,thecalibrationhasnished.ClickCloseto
closethedialog.
10Observethenumericalpresentationofthesensordepth,andverifythatitreads1m.
11Retrievethesensorfromthewater.
Howtoresetthesensorcounter
TheBottomContactandCatchsensorshavetimers.Theseareusedtokeeptrackof
howlongthesensorshavebeenactivated.
1ClicktheOperationiconundertheMainmenutoopentheOperationsub-menu.
Operationmenuonpage110
2ClicktheResetCountersbutton.
Howtosmoothoutthebottomandcatchreadings
TheCatch/BottomFilterisusedtorestrictthechangeofstatefromthePICatchandPI
BottomContactsensorstoreducejitterinthepresentation.
SettoLightltering,thechangeinstatusmustlastandremainstableforatleastt
w o
sensortransmissionsbeforethedisplayisupdated.
WhenHeavylteringisapplied,thechangeinstatusmustlastandremainstableforat
leaste i g h t sensortransmissionsbeforetheupdatedisshownonthePI50display.
TheCatch/BottomFilterisavailablebothontheOperationmenuandintheReceiver
dialog.