Instruction manual

Simrad EK60
50
850-164696 / A
C10 - Ethernet
This is the dedicated ethernet cable which connects the
transceivers with the Processing Unit computer. With more than
one GPT in use, you must use an ethernet switch and “straight”
cables.
One “straight” ethernet cable from the Processor Unit and to
the Ethernet Switch
One “straight” ethernet cable from the Ethernet Switch to
each of the transceivers.
Note Screened ethernet (CAT5 STP) must be used.
Cable details on page 65.
C20 - Remote
This is the remote power control cable from the GPT. The cable
must be connected in parallel to each of the transceivers.
Cable details on page 72.
C21 - New line
This is an input to provide manual count of survey lines. This
functionality is however not currently supported by the EK60
software.
C22 - Event
This is an input to provide a manual vertical line annotation on
the echogram. This functionality is however not currently
supported by the EK60 software.
C23 - Synchronization
The synchronization input must be connected in parallel to each
of the transceivers.
Cable details on page 73.
C24 - Motion sensor
Use this input for an analogue motion sensor. It is only
necessary to connect this input to one of the transceivers.
Cable details on page 74.
C25 - Temperature sensor
This is an input to provide water temperature from an external
sensor. This functionality is however not currently supported by
the EK60 software.
Cable details on page 74.