INSTRUCTION, OPERATING AND MAINTENANCE MANUAL FOR MODEL 4040 P/N M4040 ECO: #XX-XXXX DATE 02/22/06 DANGER Toxic and/or flammable gases or liquids may be present in this monitoring system. Personal protective equipment may be required when servicing this instrument. Hazardous voltages exist on certain components internally which may persist for a time even after the power is turned off and disconnected. Only authorized personnel should conduct maintenance and/or servicing.
Series 4040 Copyright © 2003 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-1580.
Safety Messages Your safety and the safety of others are very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument.
Series 4040 CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE PURPOSE AND IN THE MANNER DESCRIBED IN THIS MANUAL. IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE BEHAVIOR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES. This manual provides information designed to guide you through the installation, calibration and operation of your new analyzer. Please read this manual and keep it available.
Additional Safety Information DANGER COMBUSTIBLE GAS USAGE WARNING This is a general purpose instrument designed for usage in a non-hazardous area. It is the customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist. The customer should ensure that the principles of operating of this equipment are well understood by the user.
Series 4040 system. See the procedures for purging and leak checking this instrument on the following pages. If toxic gases or other hazardous materials are introduced into the sample system, the same precautions regarding leak checking and purging apply to the sample lines and sample supply or delivery lines. WARNING: ELECTRICAL SHOCK HAZARD.
Procedure for Removal of Internal Inaccessible Shock Hazards CAUTION: SERVICING OR MAINTENANCE OF THE ANALYZER SHOULD ONLY BE DONE BY SUITABLE TRAINED PERSONNEL. TO AVOID THESE INACCESSIBLE HAZARDOUS VOLTAGES WHEN SERVICING THE SERIES 4060, PERFORM EACH OF THE FOLLOWING STEPS, IN THE ORDER GIVEN, BEFORE SERVICING BEGINS: Switch off the power to the Analyzer and remove the main (line) power cord from the Analyzer. Remove all external voltages from the connections to the alarm contacts. Wait one minute.
Series 4040 DEPENDING ON THE APPLICATION, MAY ALSO USE TOXIC GASES. IT IS THEREFORE, THE CUSTOMER'S RESPONSIBILITY TO ENSURE THAT PROPER TRAINING AND UNDERSTANDING OF THE PRINCIPLES OF OPERATION OF THIS EQUIPMENT ARE UNDERSTOOD BY THE USER. SINCE THE USE OF THIS INSTRUMENT IS BEYOND THE CONTROL OF TELEDYNE, NO RESPONSIBILITY BY TELEDYNE, ITS AFFILIATES AND AGENTS FOR DAMAGE OR INJURY RESULTING FROM MISUSE OR NEGLECT OF THIS INSTRUMENT IS IMPLIED OR ASSUMED.
Table of Contents Safety Messages .......................................................................... iii Additional Safety Information ..................................................... v Procedure for Removal of Internal Inaccessible Shock Hazards vii List of Figures.............................................................................. xii List of Tables .............................................................................. xiii Introduction ......................................
Series 4040 3.3.3.1 Primary Input Power 3.3.3.2 Fuse Installation 3.3.3.3 50-Pin Equipment Interface Connector 3.3.3.4 Analog Output 3.3.3.5 Alarm Relays 3.3.3.6 Digital Remote Cal Inputs 3.3.3.7 Range ID Relays 3.3.3.8 Network I/O 3.3.3.9 Pin Out Table 3.3.4 RS-232 Port 3.3.5 Supporting Gases 3.3.5.2 Effluent 3.3.5.3 Sample Bypass Vent 3.3.5.4 Fuel and Air Connections 3.4 Placing the System in Operation 3.5 Activating the Support Gases 3.5.1 Air 3.5.2 Carrier Gas 3.5.3 Span Gas 3.5.4 Fuel 3.
4.5.2 Expanded menus 4.6 Setting up an AUTO-CAL 4.7 Adjusting the GAIN 4.8 Password Protection 4.9 Logging Out 4.10 The Model Screen 4.11 System Self-Diagnostic Test 4.12 Span 4.13 The Alarms Function 4.14 The Range Function 4.15 Changing Stream 4.16 Analog output Adjustment 4.17 Standby 4.18 Advanced User Functions 4.18.1 Timing 49 53 53 54 57 58 58 59 60 62 63 63 64 65 65 Appendix ...................................................................................... 67 A.
Series 4040 List of Figures Figure 1-1: 19” Rack Piping Diagram-Model 4060 ..................... 22 Figure 1-2: Flame Ionization Cell................................................... 24 Figure 3-1: Gas Connections......................................................... 27 Figure 3-2: Equipment Interface Connector Pin Arrangement....... 29 Figure 4-1: Front Panel View of Regulator and Gages..................
List of Tables Table 3-1: Analog Output Connections ......................................... 30 Table 3-4: Alarm Relay Contact Pins ............................................ 31 Table 3-5: Remote Calibration Connections.................................. 32 Table 3-6: Range ID Relay Connections ....................................... 33 Table 3-7: Pin out of 50 pin D-Sub Connector............................... 33 Table 3-8: Commands via RS-232 Input .......................................
Series 4040 xiv Teledyne Analytical Instruments
Introduction Introduction Teledyne Analytical Instruments Series 4040, Methane/ Nonmethane Analyzer, is designed to measure the quantity of variety of hydrocarbons present in the sample gas. The Analyzer is a microprocessor controlled digital instrument based on Teledyne’s highly successful Model 402R series analog Total Hydrocarbon Analyzer, coupled with a gas separation Column and a switching valve.
Series 4040 1.2 Principle of Operation The sample is brought to a heated compartment (50 - 80° C per application) and into a sample loop and a 10-port, 2 position (Position A & B) switching valve. In valve Position B the carrier gas, nitrogen, pushes the sample into a Column to separate methane from the remaining hydrocarbons. The eluted gases are analyzed for methane in the FID detector.
Operational Theory Series 4040 Operational Theory 2.1 Introduction The sample is brought to sample loop and a 10-port switching valve kept inside a heated (Temperature: 50° - 80° C per application) SAMPLE Chamber. Using a carrier gas and a microprocessor actuated switching valve, a fixed volume of sample collected in the Sample Loop, is pushed into the Column maintained at a constant temperature. The eluted gas is analyzed for methane in a FID detector.
Operational Theory The Series 4040 Analyzer is composed of three subsystems: 1. Sample System 2. Detector Cell 3. Electronic Signal Processing, Display and Control 2.2 Sample System The Analyzer contains three (3) separate isothermal chambers controlled by individual PID temperature controllers, viewed just behind the Front Panel. The three chambers, ‘SAMPLE’, ‘FID’, and ‘COLUMN’ are described below. The Analyzer consists of a SAMPLE Chamber containing the 10port switching valve and 2 sample loops.
Teledyne Analytical Instruments SAMPLE LOOP #2 10-PORT VALVE 2 POSITION SAMPLE LOOP #2 F.I.D. SENSOR COLUMN CHAMBER (70 C TYP) SENSOR VENT 7 6 5 4 8 2 3 1 6 9 10 7 4 5 8 2 3 1 P. REG. 3 9 10 R3 AIR IN SAMPLE LOOP #1 SAMPLE LOOP #1 IN AIR IN FUEL (H 2) (POSITION A) NMH MODE (POSITION B) CH4 MODE R2 IN REG. 1 P. REG. 1 FLOW CONTROL SAMPLE CHAMBER (50 - 80 C PER APPLICATION) R1 CARRIER FID CHAMBER (50 - 60 C TYP) IN P. REG. 2 FUEL SV1 N.C.
Operational Theory 2.2.2 Gas Flow Control System The Piping diagram is shown in Fig. 2-1. A 10 port switching valve is used to control and direct gas flows including sampling, back flush, and carrier gas. The fixed volume sample loop ensures the same volume of sample injection in the Column every cycle. 2.2.3 Fuel and Blanket Air Systems Stable flow is achieved by maintaining a constant pressure across restrictors upstream from the cell.
Operational Theory Series 4040 The anode-igniter, as its name implies, serves two functions. When relay K2 at PCB part number B74671 is energized, the coil becomes an electrical heating element that glows red-hot and ignites the hydrogen fuel. When relay K2 at B74671 is de-energized, the coil is connected to the +260 volt DC terminal of the anode-flame guard power supply PC board.
Operational Theory The electrode section of the cell may be removed for inspection by turning off the power, disconnecting the electrode lead plug, and removing the screws, which retain the electrode assembly in the sensor body. 2.3.1 Electrometer-Amplifier The collector cable is coupled directly to a coaxial fitting located on the electrometer-amplifier PC board. The PC board is located on the side panel next to but outside of the isothermal chamber.
Operational Theory Series 4040 If automatic ignition fails five times, there will be a message that reports this, and the flame can be manually ignited by pressing simultaneously the Up and Down key.
Operational Theory 24 Teledyne Analytical Instruments
Installation Installation Installation of the Series 4040 Analyzer includes: 1. Unpacking 2. Mounting 3. Gas connections 4. Electrical connections 5. Testing the system. 3.1 Unpacking the Analyzer Although the Analyzer is shipped with all the materials you need to install and prepare the system for operation. Carefully unpack the Analyzer and inspect it for damage. Immediately report any damage or shortages to the shipping agent. 3.
Installation Series 4040 Regardless of configuration, the analyzer/system must be installed on a level surface with sufficient space allocated on either side for personnel and test equipment access. Subject to the foregoing, the Analyzer/system should be placed as close to the sample point as is possible. All pertinent dimensions, connecting points, and piping details can be found in the drawings section as part of the outline, input-output, and piping diagrams.
Installation 3.3.2 Gas Connections The analyzer gas connection diagram identifies the various gas connection points as to function and location. Figure 3-1 shows the gas connection points for instruments with AutoCal module. SPAN GAS FLOW RESTRICTOR OPTIONAL SAMPLE-SPAN MANIFOLD Figure 3-1: Gas Connections Gas connections to the instrument are made at the 1/8”or 1/4” stainless steel tube fittings provided on the rear panel.
Installation Series 4040 All sample, calibration, and supporting gas lines, which deliver gas to the analyzer, must be decontaminated before connection; vent lines do not. When connecting the various gas lines to the system, be absolutely certain that no “dead ends” are left; that is, no unused branch lines should be left capped off, where pockets might form of material that is not representative of the current contents of the line, or which might keep contaminants from being purged out of the system.
Installation 3.3.3.1 PRIMARY INPUT POWER The power cord receptacle and fuse block are located in the same assembly. Insert the power cord into the power cord receptacle. CAUTION: POWER IS APPLIED TO THE INSTRUMENT'S CIRCUITRY AS LONG AS THE INSTRUMENT IS CONNECTED TO THE POWER SOURCE. The standard power supply requires 110 VAC, 50/60 Hz or 220 VAC, 50/60 Hz (optional) power. 3.3.3.2 FUSE INSTALLATION The fuse block, at the right of the power cord receptacle, accepts US or European size fuses.
Installation Series 4040 % of Range for THC (THC – Total Hydrocarbon is the summation of the CH4 and NMH values). Waveform. Waveform is the FID detection level updated 10 times each second. The output is compared against the maximum input for the Analog to Digital Conversion hardware and scaled to a 0-1VDC (or 4-20mA DC) output. Note: The Range setting applies to the CH4, NMH, and THC values.
Installation Threshold Alarm 2: Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below threshold). Configured as failsafe or non-failsafe Configured as non-latching Can be configured out (defeated). Can be set to trigger by levels of CH4, NMH, or THC. System Alarm: Actuates when power is removed from analyzer. It is triggered if 4040 fails to ignite after five times. Note: (Reset by pressing the STANDBY button to remove power.
Installation Series 4040 3.3.3.6 DIGITAL REMOTE CAL INPUTS The digital remote calibration input accepts 0 V (off) or 24 V dc (on) for remote control of calibration. (See Remote Calibration Protocol below.) See Table 3-5 for pin connections. Span: Floating input. A 5–24 V input across the + and – pins puts the analyzer into the Span mode. Either side may be grounded at the source of the signal. A 0–1 volt across the terminals allows Span mode to terminate when done.
Installation 3.3.3.7 RANGE ID RELAYS There are three dedicated Range ID relay contacts. They are assigned to relays in ascending order—Low range is assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and High range is assigned to Range 3 ID. Table 3-6 lists the pin connections. Contacts are normally open, and they close when 4040 switches to that particular range.
Installation 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 34 Series 4040 Remote Span + Remote Span - Span Solenoid Return Span Solenoid Hot Range 3 Contact Range 3 Contact Alarm 3 C Contact Range 1 Contact Range 2 Contact - Output 0-1 v (Channel 2) + Output 0-1 v (Channel 1) Alarm 1 C Contact Exhaust Solenoid Hot Sample Solenoid Hot Range 4 Contact/ not used Range 4 Contact/not used Alarm 3 NC Contact Alarm 3 NO Contact Ran
Installation 3.3.4 RS-232 Port The digital signal output is a standard RS-232 serial communications port used to connect the analyzer to a computer, terminal, or other digital device. It requires a standard 9-pin D connector. Input: The input functions using RS-232 that have been implemented to date are described in Table 3-7. Table 3-8: Commands via RS-232 Input Command Description AS Immediately starts an autospan.
Installation CAUTION: Series 4040 UNDER NO CIRCUMSTANCES SHOULD YOU EMPLOY A REGULATOR THAT IS NOT EQUIPPED WITH A METALLIC DIAPHRAGM ANYWHERE IN THE SYSTEM. The regulators should be inspected prior to installation to be sure that they are oil-free. Failure to comply with these directives will result in a constant drift in analyzer output, as organic compounds will outgas into the plumbing system at a rate that is related to the ambient temperature.
Installation 3.3.5.2 EFFLUENT All the gases introduced into the detection cell vent from one fitting at the rear of the analyzer. TAI recommends that the cell be permitted to vent directly to the atmosphere wherever possible. If a vent line is required, the installation must include a drop-out pot to collect the water that is formed by the burning of the hydrogen fuel. The vent line must be constructed so that water and dirt cannot collect in it. 3.3.5.
Installation Series 4040 3.5 Activating the Support Gases 3.5.1 Air 1. Set the air tank regulator to 30 psig. 2. Adjust the Analyzer Air regulator until the air pressure gauge reads the recommended air pressure of 9 psig. After the air is flowing through the sensor and warm-up time has been completed, activate the following gases: 3.5.
Installation Figure 3-3: Front Panel View of Regulator and Gages 3.6 Flame Ignition Observe that after warm up count down timer reaches zero (timer to preheat the sensor), the amber heater lamp is blinking (indicating that the temperature controller is maintaining the temperature setpoint) and the red flame failure lamp is on. See Figure 4-1. The Series 4040 will automatically attempt a flame ignition sequence following the warm up period, which has been preset at the factory.
Installation Series 4040 display: flame failure, check air, fuel and the flame failure LED will be on. 3. Open the cylinder supply valve and re-ignite the flame by pushing the up and down buttons simultaneously. 3.6.2 Ignition and/or Flame Guard Circuit Failure If the flame ignition or guard circuits do not operate as described in the above two sections, set the instrument fuel regulator to the recommended pressure. 3.6.3 Sample Pump A built-in Sample Pump is provided to draw Sample Gas from ambient air.
Installation Teledyne Analytical Instruments 41
Operation 42 Series 4040 Teledyne Analytical Instruments
Operation Operation 4.1 Analyzer Operation Although the Model 4040 has been programmed for your application at the factory, it can be further configured at the operator level. Depending on the specifics of the application, this might include all or a set of the following procedures: 1.
Operation Series 4040 4.2 Default Parameters The versatility of this analyzer usually results in significant changes being made to parameters over the course of time to better suit a particular application. Occasionally processes change requiring alteration to alarms, ranges etc. At some time, it may be beneficial to reset the analyzer to the default conditions as it was when shipped from the factory.
Operation Screen Modes: Screen Readout: text using all caps. Example: this is the ENTER key. Times New Roman 12 pt. italic. Example: Analysis Mode or Setup Mode. Arial Narrow, 12 pt bold. Example: AUTOCAL or Zero in 12 days. 4.4 Navigation and Data Entry Note: All menus time out after 15 (fifteen) seconds elapse with no button presses, the analyzer returns to the ANALYSIS SCREEN, if parameters have been altered but not saved (with the ENTER key), the alterations are lost.
Operation Series 4040 - Dynamic (real time) FID input level - Time spent in mode - Range - Activity indicator LINE 2 display mode 2 data: - Current analysis mode - Analysis TOTAL AREA from most recent pass of current mode. - Time spent in mode - Range - Activity Indicator Note In display mode 2, if the ESCAPE key is held down, then the TOTAL AREA for the most recent pass of the other mode is temporarily displayed (if current is CH4, then NMH TOTAL AREA is displayed). 3.
Operation 5. When the selected option is a function on the MAIN MENU or any SUB MENU screen (the menu name appears with a flashing arrow next to it), the ESCAPE key is used to return to the parent menu finally returning to the ANALYSIS SCREEN the function. If the selected option is a modifiable item (value is flashing), The ESCAPE key is used to escape to the parent menu without saving the value. The value will revert to the original before modification (if any). 4.
Operation Series 4040 - Range - Activity Indicator The MAIN MENU and SUB MENU screens have a flashing arrow on the left side indicating which item is being pointed to. MODIFIABLE ITEM screens have a flashing value somewhere in the line. 4.5.1 Main Menus The MAIN MENU consists of 13 functions you can use to customize and check the operation of the analyzer. They are listed here with brief descriptions: 1. AUTO-CAL: Used to define and/or start an automatic calibration sequence.
Operation 12. TIMING: Low level application specific timing parameters. Factory set. Do not change these. 13. STANDBY: Removes power to outputs and displays, but maintains power to internal circuitry. Any function can be selected at anytime. Just scroll through the MAIN MENU with the DOWN/UP keys to the appropriate function, and ENTER it. The analyzer will immediately start that function, unless password restrictions have been assigned. (Password assignment is explained in Section 4.6.10). 4.5.
Operation Series 4040 Reenter New Password 'A' 'A' 'A' --------------------------------------------Password Mismatch ----------------------Password Changed! ----------------------LOGOUT ----------------------Restrict Access? ENTER-YES ESCAPE-NO ----------------------MODEL ----------------------4040 HC Monitor V2.
Operation DAC A - GOOD DAC B - GOOD ----------------------SPAN ----------------------CH4 Span 2: Yes NMH Span 2: Yes CH4 Sp: 8.00 ppm NMH Sp: 2.00 ppm Span Begin --------------------------------------------Begin Span? ENTER-YES ESCAPE-NO ----------------------ALARMS ----------------------AL-1 DEFEATED AL-1 HIGH AL-1 NON-FAILSAFE AL-1 NON-LATCHING AL-1 20.00 ppm AL-1 SOURCE CH4 AL-2 DEFEATED AL-2 HIGH AL-2 NON-FAILSAFE AL-2 NON-LATCHING AL-2 50.
Operation Series 4040 R1: 0-10.00ppm R2: 0-100.00ppm R3: 0-1000.
Operation ----------------------STANDBY Switch to Sleep Mode ENTER-YES ESCAPE-NO notification if logged out ----------------------Access Restricted Re-enter Password ----------------------- 4.6 Setting up an AUTO-CAL AUTOCAL is not available for standard configurations of the 4040. 4.7 Adjusting the GAIN The output of the FID is conditioned by an amplification circuit with 3 user-adjustable settings: 0) x1 1) x10 2) x100 The resulting signal is converted to a number.
Operation Series 4040 To setup the GAIN: 1) From the Analysis Screen, press an arrow key to enter the Setup Mode. The VFD will display the first 2 lines of functions available. ________________________________________________ -> GAIN ADJUST PASSWORD ________________________________________________ <▲▼> If the arrow is not adjacent to the GAIN ADJUST menu item, use the ▲▼ keys to move the arrow to the proper position. Press to activate the function and move you to the next screen.
Operation Note however, that the instrument can still be used for analysis without entering the password. To defeat security the password must be entered. Note: If you use password security, it is strongly advised to keep a copy of the password in a separate, safe location. If the password is lost and security is enabled, the analyzer must be cold-booted. Cold booting invalidates parameters necessary for correct operation so they must be re-entered.
Operation Series 4040 The last enters the password. Steps back to the previously entered letter or, if on the first letter, returns to the MAIN MENU. If the correct password has been entered the 4040 now allows access to all the SUBMENU items. The 4040 next presents the CHANGE PASSWORD screen.
Operation Reenter a new Password 'A' 'A' 'A' ________________________________________________ Note that the new password has been replaced by 'A' as placeholder characters. This is to ensure that the wrong password is not mistakenly entered. <▲▼> Use the ▲▼ keys to reenter the letters of the new password.
Operation Series 4040 Pressing will logout the current user and protect the system from unauthorized changes to parameters. If user has LOGGED OUT, then only the following SUB MENU procedures may be executed: 17. PASSWORD 18. LOGOUT (which only repeats the logout) 19. MODEL Note however, that the instrument can still be used for analysis without entering the password. To defeat security the password must be entered. 4.
Operation Start the diagnostic testing routine by pressing or cancel out by pressing . If you pressed the self-test routine will begin and after a few moments the results will appear onscreen. The module is functioning properly if it is followed by OK otherwise it is followed by BAD. In this case, please contact TAI Customer Service for an explanation of the problem. To return the analyzer to the MAIN MENU, press after the results screen.
Operation Series 4040 Pressing enters the SPAN mode in the NMH mode first. The first pass in the NMH mode only purges the sample gas from the column (no span calculation is performed this time around). Then the CH4 is span calibrated followed by the second pass of the NMH mode, where the span calculation is performed. 4.13 The Alarms Function The Model 4040 is equipped with two alarms and a system failure alarm relay.
Operation concentration. If an active (not defeated) alarm has been set to HIGH at 20.0 ppb, and the analysis concentration is above that level, then the 'alarm condition' is occurring. DEFEATED: If an alarm is defeated, its relay is de-energized, regardless of failsafe condition. A defeated alarm does not react to a transition over its trip point in either direction.
Operation Series 4040 SOURCE: The alarm can react to measurement levels of CH4, NMH, or THC (the sum of the previous 2). CAUTION: IT IS NOT GOOD PRACTICE TO SILENCE AN EXISTING ALARM BY SETTING THE ALARM ATTRIBUTE TO ‘DEFEAT”. THE ALARM WILL NOT AUTOMATICALLY RETURN TO “ACTIVE” STATUS. IT MUST BE RESET BY THE OPERATOR BY PRESSING THE ESCAPE BUTTON. IF IT IS NOT RESET, YOUR PROCESS WILL BE RUNNING WITHOUT SAFEGUARDS THIS INSTRUMENT IS DESIGNED TO PROVIDE. 4.
Operation RANGE LIMITS: Ranges may be set as low as 50.0 ppb and as high as 1000.0 ppm. Range 3 can only be set at a value greater than Range 2. Range 2 can only be set at a value greater than Range 1. 4.15 Changing Stream Affects the relay on the rear panel that controls which gas is allowed to flow to the column and detector. When the instrument has electronically controlled calibration valves, the gas stream can be switched here.
Operation Series 4040 Use the same procedure to set the gain of the instrument. After the last digit is entered, the final press will accept the gain value. The values entered do not have explicit units of current or voltage. Adjustment should be gauged using a voltmeter or current meter attached to the analog output.
Operation To exit STANDBY, scroll again to the STANDBY function and press again. 4.18 Advanced User Functions The Model 4040 provides additional functions for tailoring the instrument to your specific application. These functions include: 20. Timing 4.18.1 Timing Timing controls the parameters of the valve and data acquisition. It is specific to the setup of the instrument and must be determined at the factory or TAI trained personnel. For further information please contact TAI customer service.
Operation 66 Series 4040 Teledyne Analytical Instruments
Appendix Appendix A.1 Specifications and Initial Settings: Range(s): Power Requirements: Signal Output: Two analog channels: - Channel 1: 0-1 VDC and isolated 4-20 mADC - Channel 2 0-1 VDC and isolated 4-20 mADC User configures what each channel tracks: CH4, non-CH4, Total HC, and Peak traces. Alarms: Two Concentration Alarm ‘C’ Type relay contacts. User configurable.
Appendix A.2 Series 4040 Recommended Spare Parts List Qty. P/N Description 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 B74671 B74677 B30927 C75825A A46128 B-13207 C62371A B74674A C65507A B74677A L79 F77 L156 L154 A33748 V825 CP2408 PC board, Flame guard & anode power supply PC board, Electrometer-amplifier PC board, Temperature control Micro-processor PC board Coaxial cable Sensor Assy. Display PCB Assy. Interface to Motherboard PCB Assy. Backpanel/Power Supply PCB Assy. Electrometer PCB Assy.
Appendix Send orders to: TELEDYNE INSTRUMENTS Analytical Instruments 16830 Chestnut Street City of Industry, CA 91749-1580 Telephone: (626) 934-1500 Fax: (626) 961-2538 Web: www.teledyne-ai.com or your local representative. Email: ask_tai@teledyne.com A.