User guide

76 Part # 11313539 Rev. B
Troubleshooting
Systemdisplays“Cannotinitializecameraframegrabbererror
Beadimagesareblurred
Redtogreenratiosuddenlychanges
Cause Resolution
Cameracableonbackof
scannerisloose.
Tightenconnection,thenstartthescannerandrestart
theiScanControlSoftware.Youmayneedtopower
cyclethescannerand/orPCseveraltimesbeforethe
framegrabberisabletoinitializesuccessfully.
Cause Resolution
Autotiltvaluesareoutof
range,slideisnot
seated
properly,oraproblem
existswiththezcontrol
board.
RemovetheBeadChipcarrierfromtheiScanReader
trayandreseattheBeadChips.Ensurethattheylie
completelyflatwithintheslotsonthecarrier.Power
cyclethescannerandrestarttheiScanControlSoftware.
Reloadthecarrierandattempt
toscanagain.Iftheerror
stilloccurs,scheduleafieldservic ecalltoexaminethe
scannerhardware.
Cause Resolution
Redorgreenlaserisnear
itsendoflife.
ContactIlluminaTechnicalSupportandscheduleafield
servicecall.