Auto-Alignment Issues
Auto-Alignment Issues
Unabletosuccessfullycompleteauto‐alignment
Abletoauto‐align,butunabletoscansectionssuccessfully
Cause Resolution
BeadChipsnotproperly
seatediniScanReader
tray.
EjectthetrayandremovetheBeadChips.Reloadthe
BeadChipsandensureproperseatingflatonthecarrier
tray.EnsurethattheBeadChipissecurelyandproperly
snappedontothetray.Itsnotchesshouldmatchthe
postsbuiltintothecarriertray,and
thelongedgeofthe
carrierclosesttotheiScanReadershouldpushdown
thesensorbutton.Ensurethatthebackofthearraysare
freeofcoating,andcleanthemiftheyaredirty.
Carriertraynotproperly
seatediniScanReader
tray.
Ejectthetray.Removethecarrierfrom
thetray.Replace
thecarrierintheiSca nReadertrayandensureproper
seating.
Sectionsmaybebrokenor
dirty.
EjecttheBeadChipsandvisuallyinspectsectionsto
ensurethatnonearebrokenordirty.Followlab
protocolsforcleaningsections,andreportbroken
sectionstotheresponsiblescientist.Tryscanningthe
BeadChipagain.
Lasersmaynolongerbe
withinfactory
specifications.
TheiScanReadermayneedtoberecalibrated.Please
contactIlluminaTechnicalSupporttorebalancelasers.
Optical‐mechanicalfailure. CallIlluminaTechnicalSupporttoinvestigatethis
further.
Cannotfindfiducials. Cleanfiducialswithanethanolwipeandrescan.
SometimestheXC4protectivecoatingon
the
BeadChipsinhibitsthelasersfromfindingfiducials.
Cameratimesout. Performapowercycle.
Systementersaninfinite
auto‐tiltloop.
Performapowercycle.
Instrumentcannotfocus. Inabilitytofocusmayresultfromusingadifferent
adaptertray.Replaceoriginaladaptertray.
Cause Resolution
ProblemsduringBeadChip
preparationmayresultin
somesectionsappearing
too
dimtoperform
registrationandbead
intensityextraction.
Contacttheresponsiblescientisttodiscussthehistoryof
theassaypreparationfortheBeadChipsandthe
scientist’sexpectation.Evaluatetheimageoutputof
nearbyimagesusingimages storedintheworkspacefor
thisBeadChip.Ifthisdoesnotresolvetheissue,contact
IlluminaTechnicalSupportforwaystocompensate.
BeadChipshavebeen
exposedtoadverse
environmentalconditions.
Evaluatetheeffectoftheenvironmentonsignalquality
pertainingtotheBeadChipsections.