15.1.3. Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the Zen-Olite™ Portable Oxygen Concentrator
The Zen-Olite™ Portable Oxygen Concentrator is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The concentrator user can
help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the concentrator
as recommended below, according to the maximum output power of the communica-
tions equipment.
Rated max. output
power of transmit-
ter (W)
Separation distance (m) according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800
MHz
d = 1.2 √P
800 MHz to 2.5
GHz
d = 2.3 √P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance (d) in meters (m) can be estimated using the equation applicable
to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Important: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
Important: These guidelines may not apply in all situations. Electromagnetic propaga-
tion is a ected by the absorption and refl ection from structures, objects, and people.
15.2. FCC WARNING STATEMENT FOR MODEL RS-00600C
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including interference that may
cause undesired operation.
This equipment complies with FCC radiation exposure limits set forth for an uncon-
trolled environment when used in the supplied carry bag or when used in a non-me-
tallic accessory that maintains a minimum 8mm separation distance from the side. End
users must follow the specifi c operating instructions for satisfying RF exposure compli-
ance. This transmitter must not be co-located or operating in conjunction with any other
antenna or transmitter.