User's Manual

EN
17/316
16/316
IMPORTANT: If an air leak is suspected, leaks can be detected with a solution of soap
and water applied to the cannula-concentrator connection point and looking for bub-
bles.
The current setting and power source (external power or battery; battery icon also shows
approximate level of charge remaining) are shown on the display screen as shown in Fig.
7.
7.5. ECO Mode
While the Setting screen is shown, press and hold the Battery button for approximately
four seconds until it changes from “Pulse XX” to “Eco Mode”. Press and hold the Battery
button again to return to Pulse Mode. At startup, the device will always default to stand-
ard Pulse Mode.
In standard Pulse Mode, the device will give you the same amount of oxygen every
breath, regardless of your breath rate. This can consume more battery power at higher
breath rates. In Eco Mode, the device will deliver a fi xed volume of oxygen per minute
regardless of breath rate, and will give an extended battery duration.
7.6. Battery Button
The battery button allows you to check the status of the battery. Repeatedly pushing
the button will cycle through all the information.
First, the gauge information will be shown
or .
Next
the number of charge cycles will be shown or
.
After the third push of the battery button, the display will alternate back to the main
screen showing the current fl ow setting. It will also automatically exit the battery status
menu and go back to the main fl ow setting display after approximately 15 seconds of no
buttons being pushed.
Fig. 7
Zen-Olite™ operating on battery
power.
Zen-Olite™ running on external
power.