IMPORTANT: Improper cannula placement may result in the
device being unable to detect all respiratory e orts of the
patient. Ensure cannula is connected securely and it has
been fully inserted.
7.2. Turning On
To turn the device on, press the power button .
The concentrator will chirp and the green, yellow, and red
LEDs will fl ash once, while the screen displays the device
name.
Red LED - indicates a warning danger and/or a need for urgent action
Yellow LED - indicates caution or attention required
Flashing green LED - indicates oxygen is being delivered.
IMPORTANT: No adjustments can be made until the startup
sequence is completed.
7.3. Choosing a Preferred Language
While the device is on, hold down the plus and mute
buttons together for approximately four seconds until
it says “Language:”.
Next cycle through the available languages using the plus
or minus buttons.
When the desired language is shown, press the battery but-
ton
to select. The device will change the language and
go back to the normal fl ow screen.
7.4. Adjusting Setting
IMPORTANT: After powering on Zen-Olite™, the startup sequence will take approximate-
ly 35 seconds. Specifi ed oxygen level will be reached within 2 minutes of use.
The device starts working in the previous setting.
The device will deliver a pulse of oxygen at the beginning of each of your inhalation.
Auto Mode: If no inhalation is detected for 60 seconds when in pulse mode, the “Check
Cannula” alarm will be activated and the device will automatically enter Auto-Mode and
continue to deliver oxygen at a rate of 18 breaths per minute. When an inhalation is
detected, the device will clear the “Check Cannula” alarm and exit Auto-Mode.
Pulse mode of operation can be adjusted from 1.0 to 5.0 in 0.5 increments with the
and buttons.
Red LED - indicates a warning danger and/or a need for urgent action
Flashing green LED - indicates oxygen is being delivered.
Cannula
port
Fig. 5
Fig. 6