3. Select the 4W function on the SM2040, Autorange. Set the calibrator to 0 . Be certain that the calibrator is
set to external sense ("EX SNS" on the Fluke 5700A or “4-Wire Comp” on the 5520A). Allow the SM2040 to
settle for a few seconds, and perform the Relative function.
4. Apply the following Resistance values to the V, + & - terminals. Check to see that the displayed reading on
the SM2040 is within the indicated range.
Table 9-4 Resistance Test, 4-wire
Step Range Input Minimum Reading Maximum Reading
1
33 [1] 0 -2 m 2 m
1
33 [1] 10 9.9972 10.0028
1
330 0 -6 m 6 m
2
330 100 99.987 100.013
3
3.3 k 0 -33 m 33 m
4
3.3 k 1 k 0.999917 k 1.000083 k
5
33 k 0 -350 m 350 m
5
33 k 10 k 9.99905 k 10.00095 k
5
330 k 0 -5 5
6
330 k 100 k 99.986 k 100.014 k
[1] SM2044 only.
Note: The use of 4-wire Ohms for resistance values above 300 k is not recommended.
6.5 AC Voltage Test
The following procedure may be used to verify the accuracy of the ACV function:
1. If you have not done so, install the SM2040 and place the covers back on to the computer. Ensure that the
computer has been on for at least one-half hour, with the covers on, before conducting this test.
2. Apply the following AC voltages to the V, + & - terminals. Check to see that the displayed reading on the
SMX2040 is within the indicated readings range.