Operator`s manual
Table Of Contents
- 1.0 Introduction
- 2.0 Specifications
- 2.1 DC Voltage Measurement
- 2.2 DC Current Measurement
- 2.3 Resistance Measurements
- 2.4 AC Voltage Measurements
- 2.5 AC Current Measurement, True RMS
- 2.6 Leakage Measurement (SMU2064)
- 2.7 RTD Temperature Measurement
- 2.8 Thermocouple Temperature Measurement
- 2.9 Additional Component Measurement Capability
- 2.10 Time Measurements
- 2.11 Trigger Functions
- 2.12 Measurement Times
- 2.12.1 Measurement Apertures and Read Interval
- 2.12.2 Range and Function Transition Times
- Range switching within Volts DC, using DMMSetRange()
- Range switching in Resistance (2-W or 4-W), using DMMSetRange()
- Switching between VDC and Resistance, using DMMSetFuncRange()
- Switching between Ohms and IDC, using DMMSetFuncRange()
- Switching between VDC and Capacitance, using DMMSetFuncRange()
- Switching between Ohms and Capacitance, using DMMSetFuncRange()
- Switching ranges within DC Current using DMMSetRange()
- Switching Capacitance ranges using DMMSetRange()
- 2.13 Source Functions (2064)
- 2.14 Accuracy Notes
- 2.15 Other Specifications
- 3.0 Getting Started
- 4.0 DMM Operation and Measurements Tutorial
- 4.1 Voltage Measurement
- 4.2 Current Measurements
- 4.3 Resistance Measurements
- 4.3.1 2-Wire Ohm Measurements
- 4.3.2 4-Wire Ohm Measurements
- 4.3.3 Using Offset Ohms function (SMU2064)
- 4.3.4 6-wire Guarded Resistance Measurement (SMU2064)
- 4.3.5 Extended Resistance Measurements (SMU2064)
- 4.3.6 Effects of Thermo-Voltaic Offset
- 4.3.7 Guarding High Value Resistance Measurements (SMU2064)
- 4.4 Leakage Measurements (SMU2064)
- 4.5 Anatomy of measurement timing
- 4.6 RTD Temperature Measurement (SMU2064)
- 4.7 Internal Temperature (SMU2064)
- 4.8 Diode Characterization
- 4.9 Capacitance Measurement, Charge Balance method
- 4.10 In-Circuit Capacitance Measurement (SMU2064)
- 4.11 Measuring the resistance in a series RC network (2064)
- 4.12 Inductance Measurement (SMU2064)
- 4.13 Characteristic Impedance Measurement (SMU2064)
- 4.14 Trigger Operation
- 4.15 Time and Frequency Measurements
- 4.16 Source Functions (2064)
- 4.17 Interfacing to an external device
- 4.18 Measuring Thermocouples’ Temperature
- 4.19 Auxiliary VDC inputs (2064)
- 5.0 Windows Interface
- 5.1 Distribution Files
- 5.2 Using the SMU2060 Driver With C++ or Similar Software
- 5.3 Visual Basic DMM Panel Application
- 5.4 Windows DLL Default Modes and Parameters
- 5.5 Using the SMU2060 DLL with LabWindows/CVI
- 5.6 Windows Command Language
- DMMArmAnalogTrigger
- DMMArmTrigger
- DMMBurstBuffRead
- DMMBurstRead
- DMMCalibrate
- DMMCleanRelay
- DMMClearMinMax
- DMMCloseUSB
- DMMDelayedTrigger
- DMMDisableTrimDAC
- DMMDisarmTrigger
- DMMDutyCycleStr
- DMMErrString
- DMMFrequencyStr
- DMMGetACCapsR
- DMMGetAperture
- DMMGetAverageVAC
- DMMGetBufferSize
- DMMGetBusInfo
- DMMGetCalDate
- DMMGetdB
- DMMGetdBStr
- DMMGetCJTemp
- DMMGetCounterRange
- DMMGetDeviation
- DMMGetDeviatStr
- DMMGetDevLocation
- DMMGetDiffMnMxStr
- DMMGetFuncRange
- DMMGetFunction
- DMMGetGrdVer
- DMMGetHwVer
- DMMGetHwOption
- DMMGetID
- DMMGetLowFreqVRMS
- DMMGetManDate
- DMMGetMax
- DMMGetMaxStr
- DMMGetMin
- DMMGetMinStr
- DMMGetNumDevices
- DMMGetRange
- DMMGetReadInterval
- DMMGetSourceFreq
- DMMGetStoredReading
- DMMGetSourceMode
- DMMGetTCType
- DMMGetTrigger
- DMMGetTriggerInfo
- DMMGetType
- DMMGetVer
- DMMInit
- DMMIsAutoRange
- DMMIsInitialized
- DMMIsRelative
- DMMLongTrigger
- DMMLongTrigRead
- DMMOpenCalACCaps
- DMMOpenTerminalCal
- DMMOpenUSB
- DMMOutputSync
- DMMPeriodStr
- DMMQuickInit
- DMMRead
- DMMReadBuffer
- DMMReadBufferStr
- DMMReadCJTemp
- DMMReadCrestFactor
- DMMReadDutyCycle
- DMMReadSR
- DMMReadFrequency
- DMMReadHiLoSense
- DMMReadHiSense
- DMMReadInductorQ
- DMMReadInductorR
- DMMReadLoSense
- DMMReadMeasurement
- DMMReadMedian
- DMMReadNorm
- DMMReadNsamples
- DMMReadPeakToPeak
- DMMReadPeriod
- DMMReadStr
- DMMReadTestV
- DMMReadTotalizer
- DMMReadWidth
- DMMReady
- DMMSetACCapsDelay
- DMMSetACCapsLevel
- DMMSetACVSource
- DMMSetAperture
- DMMSetAutoRange
- DMMSetBuffTrigRead
- DMMSetCapsAveSamp
- DMMSetCJTemp
- DMMSetCompThreshold
- DMMSetCounterRng
- DMMSetDCISource
- DMMSetDCVSource
- DMMSetFastRMS
- DMMSetFuncRange
- DMMSetFunction
- DMMSetInductFreq
- DMMSetOffsetOhms
- DMMSetPLC
- DMMSetPulseGen
- DMMSetRange
- DMMSetReadInterval
- DMMSetReference
- DMMSetRelative
- DMMSetRTD
- DMMSetSensorParams
- DMMSetSourceMode
- DMMSetSourceRes
- DMMSetSync
- DMMSetTCType
- DMMSetTempUnits
- DMMSetTrigPolarity
- DMMSetTrigRead
- DMMSetTrimDAC
- DMMStartTotalizer
- DMMStopTotalizer
- DMMTerminate
- DMMTrigger
- DMMTriggerBurst
- DMMUnlockCounter
- DMMWaitForTrigger
- DMMWidthStr
- 5.7 Calibration and Service Commands
- 5.8 Service Commands
- 5.9 Error Codes
- 5.10 Warning Codes
- 5.11 Parameter List
- 6.0 Maintenance
- 7.0 Warranty and Service
- 8.0 Accessories

Figure 4.25. Extending voltage range of auxiliary inputs by using external attenuator.
A further limitation of these function is that their common mode voltage is limited to ±3V relative to the
V- terminal. This means that to preserve accuracy and proper operation, neither terminals should have a
voltage higher than this value as it is measured between the respective terminal and the V- terminal. For
instance, while measuring a differential output voltage of 10mV of a load cell, which is stimulated by the
DMM internal DCV source, both terminals as at a common mode voltage of about 2.5V, which is within
the 3V limit. Actually the I+ terminal’s common mode is 2.505V. Provided the stimulus applied to the
cell would have been 10V, a common mode voltage of 5V would have resulted, exceeding the 3V limit.
In Figure 4.26 the stimulus voltage to the load cell is measured using the DMMRead function and the
output of the cell is measured using DMMReadHiLoSense function.
61 Signametrics