Operator`s manual
Table Of Contents
- 1.0 Introduction
- 2.0 Specifications
- 2.1 DC Voltage Measurement
- 2.2 DC Current Measurement
- 2.3 Resistance Measurements
- 2.4 AC Voltage Measurements
- 2.5 AC Current Measurement, True RMS
- 2.6 Leakage Measurement (SMU2064)
- 2.7 RTD Temperature Measurement
- 2.8 Thermocouple Temperature Measurement
- 2.9 Additional Component Measurement Capability
- 2.10 Time Measurements
- 2.11 Trigger Functions
- 2.12 Measurement Times
- 2.12.1 Measurement Apertures and Read Interval
- 2.12.2 Range and Function Transition Times
- Range switching within Volts DC, using DMMSetRange()
- Range switching in Resistance (2-W or 4-W), using DMMSetRange()
- Switching between VDC and Resistance, using DMMSetFuncRange()
- Switching between Ohms and IDC, using DMMSetFuncRange()
- Switching between VDC and Capacitance, using DMMSetFuncRange()
- Switching between Ohms and Capacitance, using DMMSetFuncRange()
- Switching ranges within DC Current using DMMSetRange()
- Switching Capacitance ranges using DMMSetRange()
- 2.13 Source Functions (2064)
- 2.14 Accuracy Notes
- 2.15 Other Specifications
- 3.0 Getting Started
- 4.0 DMM Operation and Measurements Tutorial
- 4.1 Voltage Measurement
- 4.2 Current Measurements
- 4.3 Resistance Measurements
- 4.3.1 2-Wire Ohm Measurements
- 4.3.2 4-Wire Ohm Measurements
- 4.3.3 Using Offset Ohms function (SMU2064)
- 4.3.4 6-wire Guarded Resistance Measurement (SMU2064)
- 4.3.5 Extended Resistance Measurements (SMU2064)
- 4.3.6 Effects of Thermo-Voltaic Offset
- 4.3.7 Guarding High Value Resistance Measurements (SMU2064)
- 4.4 Leakage Measurements (SMU2064)
- 4.5 Anatomy of measurement timing
- 4.6 RTD Temperature Measurement (SMU2064)
- 4.7 Internal Temperature (SMU2064)
- 4.8 Diode Characterization
- 4.9 Capacitance Measurement, Charge Balance method
- 4.10 In-Circuit Capacitance Measurement (SMU2064)
- 4.11 Measuring the resistance in a series RC network (2064)
- 4.12 Inductance Measurement (SMU2064)
- 4.13 Characteristic Impedance Measurement (SMU2064)
- 4.14 Trigger Operation
- 4.15 Time and Frequency Measurements
- 4.16 Source Functions (2064)
- 4.17 Interfacing to an external device
- 4.18 Measuring Thermocouples’ Temperature
- 4.19 Auxiliary VDC inputs (2064)
- 5.0 Windows Interface
- 5.1 Distribution Files
- 5.2 Using the SMU2060 Driver With C++ or Similar Software
- 5.3 Visual Basic DMM Panel Application
- 5.4 Windows DLL Default Modes and Parameters
- 5.5 Using the SMU2060 DLL with LabWindows/CVI
- 5.6 Windows Command Language
- DMMArmAnalogTrigger
- DMMArmTrigger
- DMMBurstBuffRead
- DMMBurstRead
- DMMCalibrate
- DMMCleanRelay
- DMMClearMinMax
- DMMCloseUSB
- DMMDelayedTrigger
- DMMDisableTrimDAC
- DMMDisarmTrigger
- DMMDutyCycleStr
- DMMErrString
- DMMFrequencyStr
- DMMGetACCapsR
- DMMGetAperture
- DMMGetAverageVAC
- DMMGetBufferSize
- DMMGetBusInfo
- DMMGetCalDate
- DMMGetdB
- DMMGetdBStr
- DMMGetCJTemp
- DMMGetCounterRange
- DMMGetDeviation
- DMMGetDeviatStr
- DMMGetDevLocation
- DMMGetDiffMnMxStr
- DMMGetFuncRange
- DMMGetFunction
- DMMGetGrdVer
- DMMGetHwVer
- DMMGetHwOption
- DMMGetID
- DMMGetLowFreqVRMS
- DMMGetManDate
- DMMGetMax
- DMMGetMaxStr
- DMMGetMin
- DMMGetMinStr
- DMMGetNumDevices
- DMMGetRange
- DMMGetReadInterval
- DMMGetSourceFreq
- DMMGetStoredReading
- DMMGetSourceMode
- DMMGetTCType
- DMMGetTrigger
- DMMGetTriggerInfo
- DMMGetType
- DMMGetVer
- DMMInit
- DMMIsAutoRange
- DMMIsInitialized
- DMMIsRelative
- DMMLongTrigger
- DMMLongTrigRead
- DMMOpenCalACCaps
- DMMOpenTerminalCal
- DMMOpenUSB
- DMMOutputSync
- DMMPeriodStr
- DMMQuickInit
- DMMRead
- DMMReadBuffer
- DMMReadBufferStr
- DMMReadCJTemp
- DMMReadCrestFactor
- DMMReadDutyCycle
- DMMReadSR
- DMMReadFrequency
- DMMReadHiLoSense
- DMMReadHiSense
- DMMReadInductorQ
- DMMReadInductorR
- DMMReadLoSense
- DMMReadMeasurement
- DMMReadMedian
- DMMReadNorm
- DMMReadNsamples
- DMMReadPeakToPeak
- DMMReadPeriod
- DMMReadStr
- DMMReadTestV
- DMMReadTotalizer
- DMMReadWidth
- DMMReady
- DMMSetACCapsDelay
- DMMSetACCapsLevel
- DMMSetACVSource
- DMMSetAperture
- DMMSetAutoRange
- DMMSetBuffTrigRead
- DMMSetCapsAveSamp
- DMMSetCJTemp
- DMMSetCompThreshold
- DMMSetCounterRng
- DMMSetDCISource
- DMMSetDCVSource
- DMMSetFastRMS
- DMMSetFuncRange
- DMMSetFunction
- DMMSetInductFreq
- DMMSetOffsetOhms
- DMMSetPLC
- DMMSetPulseGen
- DMMSetRange
- DMMSetReadInterval
- DMMSetReference
- DMMSetRelative
- DMMSetRTD
- DMMSetSensorParams
- DMMSetSourceMode
- DMMSetSourceRes
- DMMSetSync
- DMMSetTCType
- DMMSetTempUnits
- DMMSetTrigPolarity
- DMMSetTrigRead
- DMMSetTrimDAC
- DMMStartTotalizer
- DMMStopTotalizer
- DMMTerminate
- DMMTrigger
- DMMTriggerBurst
- DMMUnlockCounter
- DMMWaitForTrigger
- DMMWidthStr
- 5.7 Calibration and Service Commands
- 5.8 Service Commands
- 5.9 Error Codes
- 5.10 Warning Codes
- 5.11 Parameter List
- 6.0 Maintenance
- 7.0 Warranty and Service
- 8.0 Accessories

and negative widths of the pulse, as well as the num
ber of pulses to be generated. Use
DMMSetDCVSource to set the pulse amplitude. The latter can be set to a level of -10V to +10V. The
inactive (or negative) portion is always at 0V while the active (positive width) is set to the specified level.
The widths can be set between 25µs and 3s. The widths values are set in base units (i.e. 0.05 for 50ms).
To stop the generator, issue DMMDisableTrimDAC command. The pulse generator function requires
Driver version 1.60 and Microcode version 1.29 or higher.
Figure 4-21. Generating pulses is straight forward. It can be used for various test applications.
While the DMM is in this mode, it is possible to make DC voltage measurements. Using DMMRead will
measure the average DC at the generator output. DMMReadHiSense, DMMReadLoSense and
DMMReadHiLoSense will measure the lower two terminals.
4.17 Interfacing to an external device
The SMU2060 series of Digital Multimeters are designed to interface to various external devices, be it
Multiplexers or component Handlers. A complete handshake can be established with either devices
triggering the other, with a response scheme. The following section describes both, the hardware interface
and the software functions required to implement a synchronized operation.
The SMU2060 series can provide a complete handshake with external devices such as SMU4030 series
Relays Scanners/Multiplexers, a component Handlers or another SMU2060 series DMM. The interface
can be as simple as a single line. For instance, connecting the SMU4032 TrigOut line to the SMU2060
Trigger input. The SMU4032 is setup to provide a ready signal, indicating to the DMM it can take a
measurement. The interface can also be a two way handshake, where the a DMM Sync pulse steps the
Scanner to the next point in it’s Scan List, and the SMU4032 generates a ready signal after the relays are
stable, indicating to the DMM to make a measurement etc.. The SMU2060 series can accept a trigger
input from many sources, and send out s Sync pulse of variable width. The trigger input can be setup as
positive or negetive edge or level. Due to the limited current the Sync output can provide, it may be
necessary to boost it with a single NPN or PNP transistor (see Figure 22).
Signametrics 58