DMMSetDCVSource() since the value of the voltage is incrementaly set to the load. It takes about 10
iterations to reach the final voltage value. If a more accurate voltage is required, calibrate the source
resistace (Rs). Connect a resistor and set a voltage (within the operation envalop below), while
monitoring the voltage with an external DMM. Repeatedly make readings and set the DC voltage (as
above), while adjusting the Rs value using DMMSetSourceRes(). Using this function overrides the
DMMOpenTermianlCal() operation above. The nominal value for Rs is between 150 and 300.
Figure 4-20. The allowed Envalop of operation for V-source/I-measure funciton.
55 Signametrics