Operator`s manual
Table Of Contents
- 1.0 Introduction
- 2.0 Specifications
- 2.1 DC Voltage Measurement
- 2.2 DC Current Measurement
- 2.3 Resistance Measurements
- 2.4 AC Voltage Measurements
- 2.5 AC Current Measurement, True RMS
- 2.6 Leakage Measurement (SMU2064)
- 2.7 RTD Temperature Measurement
- 2.8 Thermocouple Temperature Measurement
- 2.9 Additional Component Measurement Capability
- 2.10 Time Measurements
- 2.11 Trigger Functions
- 2.12 Measurement Times
- 2.12.1 Measurement Apertures and Read Interval
- 2.12.2 Range and Function Transition Times
- Range switching within Volts DC, using DMMSetRange()
- Range switching in Resistance (2-W or 4-W), using DMMSetRange()
- Switching between VDC and Resistance, using DMMSetFuncRange()
- Switching between Ohms and IDC, using DMMSetFuncRange()
- Switching between VDC and Capacitance, using DMMSetFuncRange()
- Switching between Ohms and Capacitance, using DMMSetFuncRange()
- Switching ranges within DC Current using DMMSetRange()
- Switching Capacitance ranges using DMMSetRange()
- 2.13 Source Functions (2064)
- 2.14 Accuracy Notes
- 2.15 Other Specifications
- 3.0 Getting Started
- 4.0 DMM Operation and Measurements Tutorial
- 4.1 Voltage Measurement
- 4.2 Current Measurements
- 4.3 Resistance Measurements
- 4.3.1 2-Wire Ohm Measurements
- 4.3.2 4-Wire Ohm Measurements
- 4.3.3 Using Offset Ohms function (SMU2064)
- 4.3.4 6-wire Guarded Resistance Measurement (SMU2064)
- 4.3.5 Extended Resistance Measurements (SMU2064)
- 4.3.6 Effects of Thermo-Voltaic Offset
- 4.3.7 Guarding High Value Resistance Measurements (SMU2064)
- 4.4 Leakage Measurements (SMU2064)
- 4.5 Anatomy of measurement timing
- 4.6 RTD Temperature Measurement (SMU2064)
- 4.7 Internal Temperature (SMU2064)
- 4.8 Diode Characterization
- 4.9 Capacitance Measurement, Charge Balance method
- 4.10 In-Circuit Capacitance Measurement (SMU2064)
- 4.11 Measuring the resistance in a series RC network (2064)
- 4.12 Inductance Measurement (SMU2064)
- 4.13 Characteristic Impedance Measurement (SMU2064)
- 4.14 Trigger Operation
- 4.15 Time and Frequency Measurements
- 4.16 Source Functions (2064)
- 4.17 Interfacing to an external device
- 4.18 Measuring Thermocouples’ Temperature
- 4.19 Auxiliary VDC inputs (2064)
- 5.0 Windows Interface
- 5.1 Distribution Files
- 5.2 Using the SMU2060 Driver With C++ or Similar Software
- 5.3 Visual Basic DMM Panel Application
- 5.4 Windows DLL Default Modes and Parameters
- 5.5 Using the SMU2060 DLL with LabWindows/CVI
- 5.6 Windows Command Language
- DMMArmAnalogTrigger
- DMMArmTrigger
- DMMBurstBuffRead
- DMMBurstRead
- DMMCalibrate
- DMMCleanRelay
- DMMClearMinMax
- DMMCloseUSB
- DMMDelayedTrigger
- DMMDisableTrimDAC
- DMMDisarmTrigger
- DMMDutyCycleStr
- DMMErrString
- DMMFrequencyStr
- DMMGetACCapsR
- DMMGetAperture
- DMMGetAverageVAC
- DMMGetBufferSize
- DMMGetBusInfo
- DMMGetCalDate
- DMMGetdB
- DMMGetdBStr
- DMMGetCJTemp
- DMMGetCounterRange
- DMMGetDeviation
- DMMGetDeviatStr
- DMMGetDevLocation
- DMMGetDiffMnMxStr
- DMMGetFuncRange
- DMMGetFunction
- DMMGetGrdVer
- DMMGetHwVer
- DMMGetHwOption
- DMMGetID
- DMMGetLowFreqVRMS
- DMMGetManDate
- DMMGetMax
- DMMGetMaxStr
- DMMGetMin
- DMMGetMinStr
- DMMGetNumDevices
- DMMGetRange
- DMMGetReadInterval
- DMMGetSourceFreq
- DMMGetStoredReading
- DMMGetSourceMode
- DMMGetTCType
- DMMGetTrigger
- DMMGetTriggerInfo
- DMMGetType
- DMMGetVer
- DMMInit
- DMMIsAutoRange
- DMMIsInitialized
- DMMIsRelative
- DMMLongTrigger
- DMMLongTrigRead
- DMMOpenCalACCaps
- DMMOpenTerminalCal
- DMMOpenUSB
- DMMOutputSync
- DMMPeriodStr
- DMMQuickInit
- DMMRead
- DMMReadBuffer
- DMMReadBufferStr
- DMMReadCJTemp
- DMMReadCrestFactor
- DMMReadDutyCycle
- DMMReadSR
- DMMReadFrequency
- DMMReadHiLoSense
- DMMReadHiSense
- DMMReadInductorQ
- DMMReadInductorR
- DMMReadLoSense
- DMMReadMeasurement
- DMMReadMedian
- DMMReadNorm
- DMMReadNsamples
- DMMReadPeakToPeak
- DMMReadPeriod
- DMMReadStr
- DMMReadTestV
- DMMReadTotalizer
- DMMReadWidth
- DMMReady
- DMMSetACCapsDelay
- DMMSetACCapsLevel
- DMMSetACVSource
- DMMSetAperture
- DMMSetAutoRange
- DMMSetBuffTrigRead
- DMMSetCapsAveSamp
- DMMSetCJTemp
- DMMSetCompThreshold
- DMMSetCounterRng
- DMMSetDCISource
- DMMSetDCVSource
- DMMSetFastRMS
- DMMSetFuncRange
- DMMSetFunction
- DMMSetInductFreq
- DMMSetOffsetOhms
- DMMSetPLC
- DMMSetPulseGen
- DMMSetRange
- DMMSetReadInterval
- DMMSetReference
- DMMSetRelative
- DMMSetRTD
- DMMSetSensorParams
- DMMSetSourceMode
- DMMSetSourceRes
- DMMSetSync
- DMMSetTCType
- DMMSetTempUnits
- DMMSetTrigPolarity
- DMMSetTrigRead
- DMMSetTrimDAC
- DMMStartTotalizer
- DMMStopTotalizer
- DMMTerminate
- DMMTrigger
- DMMTriggerBurst
- DMMUnlockCounter
- DMMWaitForTrigger
- DMMWidthStr
- 5.7 Calibration and Service Commands
- 5.8 Service Commands
- 5.9 Error Codes
- 5.10 Warning Codes
- 5.11 Parameter List
- 6.0 Maintenance
- 7.0 Warranty and Service
- 8.0 Accessories

179 Signametrics
6.9 Frequency Counter Test ( SMU2064 only)
The following procedure may be used to verify the accuracy of the Frequency Counter:
1. If you have not done so, install the DMM and place the covers back on to the computer. Ensure that the
computer has been on for at least one-half hour, with the covers on, before conducting this test.
2. Select the ACV function, autorange. Turn freq on.
3. Apply the following AC voltages to the V, + & - terminals. Check to see that the displayed reading on the
SMU2064 is within the indicated range of readings.
ACV Frequency Counter Test
Step
Range Input Minimum Reading Maximum reading
1 240 mV 33 mV, 40 Hz 39.9952 Hz 40.0048 Hz
2 2.4 V 240 mV, 40 Hz 39.9952 Hz 40.0048 Hz
3 24 V 2.4 V, 40 Hz 39.9952 Hz 40.0048 Hz
4 330 V 24 V, 40 Hz 39.9952 Hz 40.0048 Hz
5 240 mV 250 mV, 100 kHz 99.996 kHz 100.004 kHz
6 24 V 25 V, 100 kHz 99.996 kHz 100.004 kHz
For ACI Frequency Counter test:
1. If you have not done so, install the DMM and place the covers back on to the computer. Ensure that the
computer has been on for at least one-half hour, with the covers on, before conducting this test.
2. Select the ACI function, autorange. Turn freq on.
3. Apply the following AC currents to the I,4 + & - terminals. Check to see that the displayed reading on the
DMM is within the tolerance appropriate for your application (e.g. 90 day or 1 year accuracy).
ACI Frequency Counter Test
Step
Range Input Counter Reading Tolerance
1 3.3 mA 330 uA, 40 Hz
2 33 mA 15 mA, 40 Hz
3 330 mA 150 mA, 40 Hz