Operator`s manual

Signametrics 54
trigger input from the DIN-7 connector and
ut make sure the trigger input is Disabled
ware trigger in the above sections, since
igger input operations. The Trigger input is
In) of the DMMsetPXITrigger() function. The DMM trigger input
y of the following lines.
Selected PXI Trigger input
4.18.2 Selecting PXI Trigger Inputs
The trigger input to the DMM is the wired-ored signal of the
ignal is connected to the DIN-7 trigger input while the
the selected PXI bus trigger. Make sure that no s
PXI trigger bus is in use. When using the DIN-7 trigger inp
(iTrigInput = 0). Read about the operation of the External Hard
that operation pertains to both, the external and the PXI tr
selected by the second parameter (iTrig
may be selected from an
iTrigIn
0 0 0 Disables PXI trigger input
0 0 1 PXI_TRIG1
0 1 0 PX RI_T IG2
0 1 1 PX RI_T IG3
1 0 0 PXI_TRIG4
1 0 1 PXI_TRIG5
1 1 0 PXI_TRIG6
1 1 1 PXI_STAR
The default, when the DMM is initialized (started) is for both Trigger input and output bto e deselected, or disabled.